8808A/SU 120V Fluke Electronics, 8808A/SU 120V Datasheet - Page 76

USB CABLE KIT SW DMM 5.5 RES

8808A/SU 120V

Manufacturer Part Number
8808A/SU 120V
Description
USB CABLE KIT SW DMM 5.5 RES
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheet

Specifications of 8808A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1069
8808A
Users Manual
4-16
Common Commands
*CLS
*ESE <value>
*ESE?
*ESR?
*IDN?
*OPC
*OPC?
*RST
Command
Table 4-8 describes common commands.
Clear Status
Event Status Enable
Event Status Enable Query
Event Status Register Query
Identification Query
Operation Complete
Command
Operation Complete Query
Reset
Name
Table 4-8. Common Commands
Clears all event registers summarized in the status
byte (except Message Available, which is cleared
only if *CLS is the first message in the command
line).
Sets Event Status Enable Register to <value>, where
<value> is an integer between 0 and 255.
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
If <value> is not between 0 and 255, an Execution
Error is generated.
EXAMPLE: Decimal 16 converts to binary 00010000,
which sets bit 4 (EXE) in ESE to 1.
Meter returns the <value> of the Event Status Enable
Register as set by the *ESE command.
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
Meter returns the <value> of the Event Status
Register and then clears it.
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
Meter returns the identification code of the Meter as
four fields separated by commas. These fields are:
Manufacturer (FLUKE); model (8808A); seven-digit
serial number; and versions of main software and
display software.
Meter sets the Operation Complete bit in the
Standard Event Status Register when parsed.
Meter places an ASCII 1 in the output queue when
parsed.
Meter performs power-up reset.