8808A/SU 120V Fluke Electronics, 8808A/SU 120V Datasheet - Page 22

USB CABLE KIT SW DMM 5.5 RES

8808A/SU 120V

Manufacturer Part Number
8808A/SU 120V
Description
USB CABLE KIT SW DMM 5.5 RES
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheet

Specifications of 8808A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1069
8808A
Users Manual
1-12
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
200 A
2 mA
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
Range
Notes:
[1]
Notes:
[1]
Input Protection ................................................ Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method ....................................... AC-coupled True RMS
Shunt Resistance.............................................. 0.01 Ω for 2 A and 10 A ranges
AC Filter Bandwidth:
Maximum Crest Factor ..................................... 3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ...... Crest Factor 1-2, 0.05 % of full scale
Range
Range
Slow ................................................................ 20 Hz – 100 kHz
Accuracy given as ± (% of reading + % of range)
Accuracy given as ± (% of reading + % of range)
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
(5-1/2 Digits)
19.9999 mA
199.999 mA
Frequency
Full-Scale
1.99999 A
10.0000 A
0.015 + 0.005
23 °C ± 5 °C
0.02 + 0.005
0.02 + 0.005
0.03 + 0.02
0.05 + 0.02
0.18 + 0.01
90 days
100 A
Accuracy
0.1 A
10 A
Slow
1 A
23 °C ± 5 °C
0.25 + 0.05
0.25 + 0.05
0.25 + 0.05
Input Characteristics
0.8 + 0.05
0.35 + 0.1
90 days
1 + 0.05
1 + 0.05
1 + 0.1
1 Ω for 20 mA and 200 mA
Crest Factor 2-3, 0.2 % of full scale
[1]
Accuracy
Accuracy
23 °C ± 5 °C
0.03 + 0.005
0.02 + 0.005
0.03 + 0.008
0.04 + 0.02
0.08 + 0.02
0.2 + 0.01
Resolution
1 year
Accuracy
Medium
100 A
10 A
1 mA
1 A
[1]
23 °C ± 5 °C
1.25 + 0.06
1.25 + 0.06
1.25 + 0.12
0.3 + 0.06
0.3 + 0.06
0.3 + 0.06
0.5 + 0.12
1 + 0. 06
1 year
100 A
10 A
1 mA
Fast
1 A
Temperature Coefficient/°C
Outside 18 – 28 °C
0.003 + 0.001
0.002 + 0.001
0.005 + 0.001
0.005 + 0.001
0.008 + 0.001
0.008 + 0.001
Outside 18 – 28 °C
Burden Voltage
Coefficient/°C
Temperature
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
<0.05 V
<0.5 V
<0.1 V
<0.5 V