8808A/SU 120V Fluke Electronics, 8808A/SU 120V Datasheet - Page 23

USB CABLE KIT SW DMM 5.5 RES

8808A/SU 120V

Manufacturer Part Number
8808A/SU 120V
Description
USB CABLE KIT SW DMM 5.5 RES
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheet

Specifications of 8808A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1069
Frequency
Continuity
Diode Test
100 mV to
750 V
Range
Notes:
[1]
[2]
Gate Time .......................................................... 131 ms
Measurement Method ....................................... AC-coupled input using the ac voltage measurement function.
Settling Considerations ................................... When measuring frequency after a dc offset voltage change, errors
Measurement Considerations ......................... To minimize measurement errors, shield inputs from external noise
Continuity Threshold........................................ 20 Ω
Test Currents .................................................... 1 mA
Response Time ................................................ 100 samples/sec with audible tone
Rate…….…………………………………………… Fast
Maxiumum Reading………………………………199.99 Ω
Resolution……….….. ……………….……………0.01 Ω
Response Time ................................................. 100 samples/sec with audible tone
Rate………………………………………………….Fast
Maxiumum Reading………………………..….…1.9999 V
Resolution….………..….. ………………………..0.1 mV
[1,2]
Input > 100 mV
Limited to 8* 10
20 kHz – 200 kHz
200 kHz – 1 MHz
2 kHz – 20 kHz
20 Hz – 2 kHz
Frequency
7
V Hz
23 °C ± 5 °C
0.01 + 0.002
0.01 + 0.002
0.01 + 0.002
0.01 + 0.004
90 days
may occur. For the most accurate measurement, wait up to 1 second
to allow input blocking RC time constant to settle.
when measuring low voltage, low frequency signals.
Accuracy
Accuracy
0.01 + 0.003
0.01 + 0.003
0.01 + 0.003
23 °C ± 5 °C
0.01 + 0.006
1 year
Introduction and Specifications
Electrical Specifications
Outside 18 – 28 °C
Coefficient/°C
Temperature
0.002 + 0.001
0.002 + 0.001
0.002 + 0.001
0.002 + 0.002
1-13