OS523/N NEWPORT ELECTRONICS, OS523/N Datasheet - Page 67

Temperature Display Panel

OS523/N

Manufacturer Part Number
OS523/N
Description
Temperature Display Panel
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS523/N

Temperature Measuring Range
-18°C To +1371°C
Temperature Tester Type
Thermometer - Infrared
External Height
167.6mm
External Width
218.4mm
Lead Free Status / RoHS Status
na
Optics Field of View
A-4
A
Appendix: How Infrared Thermometry Works
Accurate measurement of temperature via infrared means depends
strongly on the size of the object and the distance between the
thermometer and the object. All optical devices (e.g. cameras,
microscopes, infrared thermometers) have an angle of vision,
known as a field of view or FOV, within which they see all objects.
In particular, the thermometer will measure a fixed proportion of
the energy radiated by all objects within its FOV. The user must
guarantee that the distance between the thermometer and the object
is defined so that only that object fills the FOV of the instrument.
Referring to Figure A-3, Objects “X” and “Y” are within the FOV of
the thermometer. The measured temperature would fall somewhere
between the actual temperatures of the two objects. In order to
measure the temperature of Object “X” accurately, Object “Y”
would need to be removed. In order to measure the temperature of
Object “Y” accurately, the user would need to move closer to Object
“Y” until it completely filled the FOV of the thermometer.
Alternatively, the user could measure the temperature of Object “Y”
with a thermometer with a smaller FOV.
The distance-to-spot size ratio (
Thus, a
distance of 10 feet. For accurate spot size values, refer to the Field of
View diagrams shown in Figures 2-4 through 2-6.
D
S
Figure A-3. Field of View of a Thermometer
= 10 gives you approximately a 1 foot spot size at a
ANGLE
FOV
DISTANCE
D
S
) defines the field of view (FOV).
OBJECT "Y"
SPOT
SIZE
OBJECT "X"