OS523/N NEWPORT ELECTRONICS, OS523/N Datasheet - Page 32

Temperature Display Panel

OS523/N

Manufacturer Part Number
OS523/N
Description
Temperature Display Panel
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS523/N

Temperature Measuring Range
-18°C To +1371°C
Temperature Tester Type
Thermometer - Infrared
External Height
167.6mm
External Width
218.4mm
Lead Free Status / RoHS Status
na
ATC
2.3.8 Using Ambient Target Temperature Compensation
°F
(OS533, OS534, OS523, OS524)
Use the Ambient Target Temperature Compensation
(AMB)
of these conditions are required:
1.
2.
3.
4.
5.
6.
7.
To set and activate the Ambient Target Temperature
Compensation Mode:
• The target has a low emissivity.
• The ambient temperature around the target is much
Pull the trigger and press the
trigger. Set the emissivity to 1.0 (refer to Section 2.3.1).
Press and hold the
Mode (AVG) appears.
Slowly move the thermometer so that the line of sight
sweeps across the area surrounding the target. The
thermometer measures the temperature at each point on
the surrounding area.
Read the average temperature value from the upper
display and record it here __________.
Press and hold the
Temperature Display Mode (AMB) appears.
Set the AMB temperature found in Step 4 by pressing the
Press the
temperature compensation. The
on the display.
higher than the ambient temperature around the
infrared thermometer.
Display Mode when high accuracy readings under both
To disable this mode, press the
The
key or the
Using the Handheld Infrared Thermometer
icon disappears.
key to enable the ambient target
key.
NOTE
key until the Average Display
key until the Ambient
key to lock the
key again.
icon appears
2