OS523/N NEWPORT ELECTRONICS, OS523/N Datasheet - Page 29

Temperature Display Panel

OS523/N

Manufacturer Part Number
OS523/N
Description
Temperature Display Panel
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS523/N

Temperature Measuring Range
-18°C To +1371°C
Temperature Tester Type
Thermometer - Infrared
External Height
167.6mm
External Width
218.4mm
Lead Free Status / RoHS Status
na
2-14
LCK
2
2.3.6 Thermocouple Input (OS532, OS533, OS534)
°F
Using the Handheld Infrared Thermometer
The thermometer accepts thermocouple input. It displays
thermocouple temperature and the target temperature (via
infrared) simultaneously. This function provides an accurate
method of determining an unknown emissivity.
• To Determine an unknown target emissivity
LCK
1.
2.
3.
4.
5.
6.
Connect a contact thermocouple probe (Type K) to the
thermometer as shown in Figure 1-3.
Measure the object temperature using the thermocouple
probe.
Aim at the object and measure the temperature via infrared.
Press and hold the
mode (E) appears.
Set the emissivity by pressing the
the temperature reading matches the thermocouple
temperature measurement.
The thermometer now displays the correct object
emissivity.
°F
"TC--- is” is displayed when the thermocouple
input is open or out of range (0 to 1600°F).
key until the Emissivity Display
or
keys until