OS523/N NEWPORT ELECTRONICS, OS523/N Datasheet - Page 56

Temperature Display Panel

OS523/N

Manufacturer Part Number
OS523/N
Description
Temperature Display Panel
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS523/N

Temperature Measuring Range
-18°C To +1371°C
Temperature Tester Type
Thermometer - Infrared
External Height
167.6mm
External Width
218.4mm
Lead Free Status / RoHS Status
na
The temperature reading is
erratic. The thermometer
has just been moved from
one extreme temperature
to room temperature [0°C
or 50°C (32°F or 122°F)]
or vice versa.
The temperature reading
is erratic. The thermometer
has just been moved from
room temperature
(ambient temperature) to a
temperature 10°C colder
or warmer.
Problem
1.
1.
If you see an error code, either
“ERR1”, “ERR2”, or “ERR3”, record
the code and call our Customer
Service Department. Provide
Customer Service with the error
code that is displayed in the upper
left corner of the display. The
Customer Service Department
representative may ask you to return
the thermometer to the factory.
The display will go back to the
Real Time Mode (Emissivity
Display Mode).
c.
The thermometer has to stabilize
before taking temperature
measurements. It takes up to 40
minutes for the thermometer to
stabilize.
The thermometer has to stabilize
before taking temperature
measurements. It takes up to 20
minutes for the thermometer to
stabilize.
After running the diagnostic
program, press the
unlock the trigger.
Troubleshooting Guide
Solution
key to
6