U3402A User’s and Service Guide
To Replace the Power Line Fuse
The power line fuse is located within the multimeter’s fuse- holder assembly
on the rear panel. The multimeter is shipped from the factory with a
power- line fuse installed (according to country of destination). See
If you determine that the fuse is faulty, replace it with one of the same size and
rating.
Table 5-1 Type of supplied fuse (according to country of destination)
1 Disconnect the power cord. Depress tab 1 and
3 Rotate the line voltage selector and reinstall
Type of fuse (time-lag, low breaking fuse)
0.25A, 250V, 5x20 mm
0.125A, 250V, 5x20 mm
2 and pull the fuse holder from rear panel.
it back, so that the correct voltage appears in
the fuse holder window.
100, 120, 220 (230), or 240 Vac
1
2
Input line voltage
100 V to 120 V
220 V to 240 V
2 Remove the line voltage selector from fuse
4 Replace fuse holder assembly in rear panel.
holder assembly.
Disassembly and Repair
Part number
A02-62-25592-3U
A02-62-25648-1U
Table 5-
1.
79