Figure 2-3 ACI RMS or DCI (mA) terminal connection and display
U3402A User’s and Service Guide
Performing Current Measurements
Measuring AC (RMS) or DC Current
1 Press
2 Power off the measured circuit.
3 Connect the red and black test leads to mA input terminal as shown in
4 Probe the test points in series with the circuit
5 Power on the measured circuit and read the display.
• Four AC current or DC
• Shunt resistance:
• Input protection:
Figure 2-
current ranges:
ACI or DCI
current source
+
3.
or
.
• Slow reading rate: 12.0000 mA, 120.000 mA, 1.20000 A
• Medium reading rate: 40.000 mA, 120.00 mA, 1.2000 A
• Fast reading rate: 40.00 mA, 120.0 mA, 1.200 A
0.1
Front panel 1.25 A, 500 V FH fuse for one terminal
to 10
in mA
for 12 mA to 1.2 A range
Operations and Features
23