U3402A-ABA AGILENT TECHNOLOGIES, U3402A-ABA Datasheet - Page 116

no-image

U3402A-ABA

Manufacturer Part Number
U3402A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3402A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3402A Dual Display Multimeter
6
Table 6-4 ACV
Rate
[1]
[2]
[3]
[4]
96
Slow
Medium
Fast
In VAC 750 V range, 787.5 V RMS is readable.
For input < 500 V RMS.
VDC and VAC are automatically set at the same range.
Specified accuracy at input > 5% of full scale.
Specifications and Characteristics
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
750 V
ac+dc
[1]
AC Voltage (True RMS, AC+DC Coupling Mode)
resolution, full scale reading, and accuracy [± (% of reading + count)]
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
Maximum
reading
119.999
1.19999
11.9999
119.999
750.00
399.99
3.9999
39.999
399.99
750.0
399.9
3.999
39.99
399.9
750
[3]
45 Hz to 10kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 45
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
Accuracy (One year; 23ºC ± 5ºC)
10 kHz to 30 kHz
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 83
1% + 43
1% + 43
1% + 43
1% + 43
1.5% + 12
1% + 7
1% + 7
1% + 7
1% + 7
U3402A User’s and Service Guide
30 kHz to 100 kHz
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
5% + 125
3% + 83
3% + 83
3% + 83
3% + 83
5% + 18
3% + 12
3% + 12
3% + 12
3% + 12
[2]
[4]