CAU T ION
Operations and Features
N O TE
Disconnect circuit power and discharge all high-voltage capacitors before testing
diodes to avoid damaging the multimeter.
To test a diode, switch the circuit power off, and remove the diode from the
circuit. Then proceed as follows:
1 Press
2 Connect the red and black test leads to the input terminal as shown in
3 Connect the red test lead to the positive terminal (anode) of the diode and
The cathode of a diode is indicated with a band.
4 Read the display.
5 Reverse the probes and measure the voltage across the diode again (refer to
• Measurement method:
• Response time:
• Gate time:
• Input protection:
Figure 2-
the black test lead to the negative terminal (cathode). Refer to
Figure 2-
• A diode is considered good if the multimeter displays OL in reverse bias
• A diode is considered shorted if the multimeter displays approximately
• A diode is considered open if the multimeter displays OL in both forward
mode.
0 V in both forward and reverse bias modes, and the multimeter beeps
continuously.
and reverse bias modes.
8.
9). Assess the diode according to the following guidelines:
. The default function is diode measurement.
0.83 mA±0.2% constant current source, open-circuit voltage
limited to <5 V
70 samples per seconds with audible tone
0.1 s or 1 period of the input signal, whichever is longer
500 V RMS on all ranges
U3402A User’s and Service Guide