6
Table 6-6 ACI
Rate
[1]
[2]
98
Slow
Medium
Fast
Typical at full scale reading and voltage across the input terminals.
Specified accuracy at input > 5% of full scale.
Specifications and Characteristics
Range
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
40.000 mA
120.00 mA
1200.0 mA
12.000 A
40.00 mA
120.0 mA
1200 mA
12.00 A
ac+dc
AC Current (True RMS, AC+DC Coupling Mode)
resolution, full scale reading, burden voltage, and accuracy [± (% of reading + count)]
Resolution
0.1 µA
1 µA
10 µA
100 µA
1 µA
10 µA
100 µA
1 mA
10 µA
100 µA
1 mA
10 mA
Maximum
reading
11.9999
119.9999
1199.99
11.9999
39.999
119.99
1199.9
11.999
39.99
119.9
1199
11.99
Burden voltage
and shunt resistor
<0.15 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
<0.5 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
<0.5 V/10
<1.5 V/10
<0.3 V/0.1
<0.6 V/0.01
[1]
U3402A User’s and Service Guide
Accuracy (one year; 23ºC ± 5ºC)
45 Hz to 2 kHz
0.5% + 100
0.5% + 100
0.5% + 100
1% + 100
0.5% + 42
0.5% + 15
0.5% + 15
1% + 15
0.5% + 7
0.5% + 4
0.5% + 4
1% + 4
2 kHz to 10 kHz
2% + 200
2% + 200
2% + 200
2% + 80
2% + 30
2% + 30
2% + 12
2% + 7
2% + 7
[2]