U3402A User’s and Service Guide
CAU T ION
DC Current Verification Test
1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in
3 Make a measurement and observe the result. Compare measurement results
Table 4-3 DC current verification test
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.
Function
DC current
input shown in
to the appropriate test limits shown in
appropriate source settling when using the Fluke 5520A.)
Reading rate
Slow
Medium
Slow
Medium
Table 4-
3.
Input
0.0000 mA
0.000 mA
0.00 mA
0.0000 A
0.000 mA
0.00 mA
0.0 mA
0.000 A
10.0000 mA
100.000 mA
1000.00 mA
10.0000 A
36.000 mA
100.00 mA
1000.0 mA
10.000 A
Table 4-
Range
12 mA
120 mA
1200 mA
12 A
40 mA
120 mA
1200 mA
12 A
12 mA
120 mA
1200 mA
12 A
40 mA
120 mA
1200 mA
12 A
3. (Be certain to allow for
Table 4-
Performance Test
Error from nominal
one year
±0.5 mA
±20.5 mA
±23 mA
±1.5 µA
±5 µA
±50 µA
±6 µA
±30 µA
±0.3 mA
±3 mA
±6.5 µA
±55 µA
±1.55 mA
±42 µA
±0.13 mA
±1.8 mA
3. Provide the
67
4