TL1550EXT Fluke, TL1550EXT Datasheet - Page 5

no-image

TL1550EXT

Manufacturer Part Number
TL1550EXT
Description
TEST LEADS
Manufacturer
Fluke
Datasheet

Specifications of TL1550EXT

Cable Length
25ft
Lead Length
25ft
Display Type
Component Test Equipment and Logic Probes
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
*Addresses pulsed transmitters and PLCs with pulses as short as 1 ms. For pressure specifications, see page 7.
Summary one year specifications for Fluke 741, 743B and 744 calibrators
General specifications for Fluke
741, 743B and 744 calibrators
Loop power: Selectable, 24 V or 28 V;
22 mA max
Environmental: 741/743 specifications
apply from +18 °C to +28 °C
Operating temperature: -10 °C to 50 °C
(typical specs to -20 °C) Pressure modules
are totally compensated and specs apply
0 °C to 50 °C
Storage temperature: -20 °C to 60 °C
Range (full scale)
Volts dc
110.000 mV dc
1.10000, 11.0000 V dc
110.000, 300.00 V dc
Volts ac
V ac, 20 to 40 Hz
V ac, 40 to 500 Hz
V ac, 500 to 1 kHz
V ac, 1 kHz to 5 kHz
mA
30.000 mA dc
110.00 mA dc
Resistance
11.000 Ω
110.00 Ω
1.1000 kΩ
11.000 kΩ
Frequency
1.00 to 109.99 Hz
110.0 to 1099.9 Hz
1.100 to 10.999 kHz
11.00 to 50.00 kHz
Temperature
10 thermocouples
11 RTDs*
Measure
(% of reading + % of full scale)
0.025 % + 0.015 %
0.025 % + 0.005 %
0.05 % + 0.005 %
0.05 % + 50 mΩ
0.05 % + 50 mΩ
0.05 % + 0.5 Ω
0.01 % + 0.015 %
0.01 % + 0.015 %
0.5 % + 5
0.1 % + 10 Ω
10 % + 20
2 % + 10 counts
2 % + 10
Accuracy
0.05 Hz
0.5 Hz
0.3 °C
0.3 °C
50 Hz
5 Hz
Safety: Complies with CAN/CSA C22.2 No.
1010.1-92, ANSI/ISA S82.01-1994, UL3111
and EN610-1:1993. 300 V, Over-voltage
Category II
Size/weight: 130 x 236 x 61 mm (5.1 x 9.3
x 2.4 in), weight 1.4 kg (3 lbs, 1 oz)
Battery: Internal battery pack NiCd, 7.2 V,
1700 mAh; NiMH (744 only) 7.2 V, 3500
mAh, typical usage, greater than 8 hours
Warranty: Three years
Simulate 22.000 mA
22.000 to 50.000 kHz
Source 22.000 mA
11.00 to 109.99 Hz
110.0 to 1099.9 Hz
1.100 to 21.999 kHz
0.00 to 0.99 Hz
Range (full scale)
110.000 mV
1.10000 V
1.1000 kΩ
11.000 kΩ
11.000 Ω
110.00 Ω
Source
(% of reading + % of full scale)
0.01 % + 0.005 %
0.01 % + 0.005 %
0.02 % + 0.03 %
0.02 % + 0.5 Ω
0.03 % + 5 Ω
0.01 % + 0.015 %
0.01 % + 20 mΩ
0.01 % + 40 mΩ
Accuracy
0.01 Hz
0.1 Hz
0.1 Hz
0.2 °C
0.1 °C
2 Hz
5 Hz
Ordering information
Fluke-741B Documenting
Process Calibrator
Fluke-743B Documenting
Process Calibrator
Fluke-744 Documenting
Process Calibrator
Each calibrator includes TL224
Industrial Test Leads (2 sets),
AC220 Test Clips (2 sets),
TP220 Test Probes (1 set),
BP7217 Battery Pack, BC7217
Battery Charger, instruction
manual, NIST traceable calibra-
tion certificate with data, three-
year warranty, serial port cable
(743B and 744 only),
DPC/TRACK Sample Version
with free PC communication
utility software (743B and 744
only). With the 744 you also get
a HART communication cable,
HART Users Manual and a NiMH
battery (instead of NiCad).