TL1550EXT Fluke, TL1550EXT Datasheet - Page 19

no-image

TL1550EXT

Manufacturer Part Number
TL1550EXT
Description
TEST LEADS
Manufacturer
Fluke
Datasheet

Specifications of TL1550EXT

Cable Length
25ft
Lead Length
25ft
Display Type
Component Test Equipment and Logic Probes
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
Specifications (18 °C to 28 °C, one year)
1 Supported pressure units; psi, in. H
2 Pressure module specifications, see page xx
inHg, mmHg, kg/cm
2
Function
Measure pressure
(internal sensor)
Over pressure 3xFS
Measure Pressure
(internal sensor)
Over pressure 2xFS
Measure pressure
(with pressure
modules, modules)
Over pressure per
pressure
module specs
Source pressure,
built-in pump
Measure mA
Loop supply
2
2
0 (4 °C), in. H
1
1
1
Range
-12 to 30 psi
(-83 to 207 kPa) scale (0.01 kPa)
-12 to 100 psi
(-83 to 690 kPa) (0.1kPa)
29 Pressure
1.0 in. H20/
0.25 kPa to
10,000 psi/
69 MPa
-11 psig to
full scale
0 to 24 mA
24 V dc
2
0 (20 °C), cm H
Resolution
0.001 psi
0.01 psi
To 0.0001 psi,
per pressure
module specs
N/A
0.001 mA
N/A
2
0 (4 °C), cm H
2
Accuracy
0.05 % full
0.05 %
full scale
To 0.05 %
of full
span, per
pressure
N/A
0.025 %
+ 1 count
± 10 %
2
0 (20 °C), bar, mbar, kPa,
Notes
Gases/liquids 717
only (non-corrosive)
Use the 718 with
non-corrosive
gases only
Zero, Min, Max, Hold,
Damp
Gases/liquids 717
only (non-corrosive)
Use the 718
with non-corrosive
gases only
Zero, Min, Max, Hold,
Damp
Media compatibility
per Pressure Module
Specs
Min, Max,
Hold, Damp
Use the 718 with
non-corrosive
gases only
Drive; 20 mA into
1000 Ω
2
Zero,
General specifications
for all Fluke 717 and
718 calibrators
Maximum voltage: 30 V
Non-operating temperature:
-40 °C to 60 °C
Operating temperature:
-10 °C to 55 °C
Relative humidity:
95 % (10 °C to 30 °C); 75 %
(30 °C to 40 °C); 45 % (40 °C to
50 °C); 35 % (50 °C to 55 °C)
Operating altitude:
3,000 m max
Shock: 1 m drop test
Vibration: Random, 2 g,
5-500 Hz
Safety: CSA C22.2 No.
1010.1:1992 (Call Fluke for
718Ex safety ratings)
EMC: EN50082-1:1992 and
EN55022:1994 Class B
Size/weight, (717 with holster
and Flex-Stand ™ ):
201mm L x 98 mm W x 52 mm D
(7.93 in L x 3.86 in W x 2.06 in
D) 600 g (21 oz)
Size/weight, (718 with holster):
216 mm L x 94 mm W x 66 mm
D (8.50 in L x 3.72 in W x
2.60 in D) 992 g (35 oz)
Power: 9 V battery ANSI/NEDA
1604 A or IEC 6LR61 9V
alkaline; two batteries in 718
Battery Life: 4 to 20 hours, typi-
cal, depending on functions used
Warranty: Three years (one year
on pressure pump in Fluke 718);
one year for 718Ex.