Summary specifications for 1502A and 1529
Ordering information
1529
1529-R
1529-T
(Specify 110V or 220V power)
2506-1529
9322
1502A-156
1502A-256
2502
2506
2508
9313
9301
9308
Inputs
Temperature range
Measurement range
Characterizations
Temperature accuracy,
typical (meter only)
Temperature resolution
Operating range
Logging intervals
Communications
Size (HxWxD)
Weight
Calibration
Thermometer
Readout
Battery Pack
Chub-E4, 2 TC and 2 PRT/Thermistor Inputs
Carrying Case, fits Tweener and 12 in. probe
Rugged Carrying Case
DC Power Option
Serial Cable Kit
Carrying Case, fits Tweener and 6 in. probe
IEEE Option
Chub-E4, 4 TC Inputs
Chub-E4, 4 PRT/Thermistor Inputs
“Tweener” PRT Thermometer (110 V)
IEEE Option
“Tweener” PRT Thermometer (220 V)
PRT / RTD
2 channels PRT/thermistor and 2 channels TC, or 4 channels PRT/thermistor, or
4 channels TC, specify when ordering; PRT/thermistor channels accept 2, 3, or
4 wires; TC inputs accept B, E, J, K, N, R, S, T, and Au-Pt TC types
–189 °C to 960 °C
(-308 °F to 1760 °F)
0 to 400 Ω
ITS-90, IEC-751
(DIN “385”),
Callendar-Van Dusen
± 0.006 °C at 0 °C
± 0.009 °C at 100 °C
0.001°
16 °C to 30 °C
0.1, 0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes
RS-232 (tweener) and IR ports included, IEEE-488 (GPIB) optional
4.0 x 7.5 x 8.2 in (102 x 191 x 208 mm)
4.5 lb (2 kg)
Accredited NIST-traceable resistance calibration and NIST-traceable voltage calibration provided
Thermistor
–50 °C to 150 °C
(-58 °F to 302 °F)
0 to 500 KΩ
Steinhart-Hart,
YSI-400
± 0.0025 °C at 0 °C
± 0.025 °C at 100 °C
0.0001°
Hart Scientific 1529
Chub-E4 Thermometer
Lab-quality accuracy on
four channels
Hart Scientific 1502A “Tweener” Thermometer
Best performance thermometer in its price range
Four channels for PRTs, ther-
mistors and thermocouples
Displays eight user-selected
data fields
Logs up to 8,000 readings
Battery provides eight hours of continuous operation
PRT readout with accuracy to ± 0.006 °C
Reads both 100 Ω and 25 Ω probes
0.0001 °C resolution
Optional battery pack available
Compatible with LogWare and MET/TEMP II software
Thermocouple
–270 °C to 1800 °C
(-454 °F to 3272 °F)
–10 to 100 mV
NIST Monograph 175, 3-point
deviation function applied to NIST
175, 6th-order polynomial
K at 600 °C ± 0.15 °C: ± 0.4 °C
T at 200 °C ± 0.1 °C:
0.01 to 0.001°
Ext. RJC:
Int. RJC
± 0.3 °C
Tweener
-200 °C to 962 °C
(-328 °F to 1764 °F)
0 to 400 Ω
ITS-90 subranges 4,
6 thru 11, IPTS-68
Callendar-Van Dusen
± 0.006 °C at 0 °C
± 0.009 °C at 100 °C
0.001 °C
2.4 x 5.6 x 7.1 in
(61 x 143 x 181 mm)
2.2 lb (1.0 kg)