DV3100 DIGIVIEW, DV3100 Datasheet - Page 73

no-image

DV3100

Manufacturer Part Number
DV3100
Description
LOGIC ANALYSER, DIGITAL, 100MHZ, 18CH
Manufacturer
DIGIVIEW
Datasheet

Specifications of DV3100

No. Of Data Channels
18
No. Of Clock Inputs
8
Frequency
100MHz
Power Consumption
2.5W
External Height
0.75"
External Width
2.8"
External Depth
4.75"
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
© 2010 TechTools
4.9
Using Link Groups
Link groups provide Synchronous Scrolling of a group of windows that display data. When
several windows are part of the same Link Group, each window will automatically update and
scroll when any member of the group changes its center time. The time reference and displayed
data of each member will "center" on the time the changing member centers on. This feature
keeps a group of windows synchronized to the same time in the captured data.
DigiView provides five link options that include Link Groups 1 through 4 and a "no-link" option for
windows that you want to keep independent.
Default link groups:
Asynchronous Signal - The list format for this signal consists of a HEX, ASCII and Status
column.
Synchronous Signal - The list format for this signal consists of a HEX and ASCII column.
The ASCII column will be blank if the packet value is greater than an 8 bit ASCII value. If a
packet is terminated prematurely when using the Enable qualifier, the HEX value will display
the letter T.
State Signal - The list format for this signal consists of a HEX and ASCII column. If a packet is
terminated prematurely when using the Enable qualifier, the HEX value will display the letter T.
Link Group 1 - Waveform View 1 is always a member and all new windows initially belong to
this group.
•CBUS - Beginning of CBUS packets
•RSRVB - Reserved Bus Address
•RSRV1 - First Block of Reserved Codes
•HS-MS - High Speed Mode Master Code
•RSRV2 - Second Block of Reserved Codes
•10-bit - 10 bit Slave Address
•PGM!R - Program Slave Address without resetting
•PGM+R - Program Slave Address and Reset Slave
•MASTR - 7 bit Master Address
•ERROR - Illegal Write of ZERO
•!DEF! - Code is Not Defined
Status:
•" " - No Errors
•F - Framing Error
•P - Parity Error
•B - Break
Navigating and Analyzing the Data
69