52
Capture Troubleshooting
DigiView User's Guide
1. DigiView (DV1-100, DV3100) Power LED
does not stay ON.
2. DigiView is only identified as a SERIAL
device by the USB subsystem.
3. One of DigiView's channels is connected to
a transitioning signal, but the capture does not
seem to show any activity or very little activity.
Symptom
(A). The USB subsystem of the PC is turning
DigiView Off due to a Bandwidth shortage.
REMEDY: Remove other USB devices that
use a port on the same hub (internal or
external).
(B). The USB subsystem of the PC is turning
DigiView Off or DigiView is resetting due to
insufficient power.
REMEDY: If you are using an external hub,
change to a POWERED external hub or
purchase one of better quality that meets USB
specifications.
(A). The USB sub-system is experiencing a
problem with corrupted or missing files.
REMEDY: Look for a driver update to the USB
hub (internal or external) or re-install its
drivers. Refresh the DigiView drivers by
re-installing the DigiView software.
(B). Operating System or computer needs
rebooted.
REMEDY: Turn off the computer and re-boot
the system.
(C). The DigiView hardware has been
damaged.
REMEDY: Call Technical Support for repair
information.
(A). The resolution of the Waveform View is
too low to see the activity.
REMEDY: Zoom In to see the details of the
activity.
(B). The signal has been defined with the
wrong channel selected.
Causes & Solutions