DV3100 DIGIVIEW, DV3100 Datasheet - Page 43

no-image

DV3100

Manufacturer Part Number
DV3100
Description
LOGIC ANALYSER, DIGITAL, 100MHZ, 18CH
Manufacturer
DIGIVIEW
Datasheet

Specifications of DV3100

No. Of Data Channels
18
No. Of Clock Inputs
8
Frequency
100MHz
Power Consumption
2.5W
External Height
0.75"
External Width
2.8"
External Depth
4.75"
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
© 2010 TechTools
button to add it as the Less Than Match. You can also just "Double-Click" the defined signal to
automatically add it. NOTE: High Level signal definitions (such as State, I2C, or etc) cannot be
added directly as a complete signal. However, the individual portions of the signal (such as the
DATA portion of a State Signal) can be added in the same manner. To see the sub-portions of a
signal, click the "expand" button in the left column next to the signal's name, then add the sub
portion in the same manner as any other signal.
For instance, if you have defined a Synchronous Serial signal (i.e. SPI) and you want to add the
:Enable portion of the signal to the Less Than match, then expand it and double-click on :Enable.
It would then appear on the right portion of the editor.
The Less Than Match type only accepts one signal at a time. After a signal has been added, the
"+" (ADD) button will be disabled (as shown above) until the current signal is removed.
Delete a Signal - Highlight a signal in the right portion of the window (or mouse-over its bit
selections) and click the "-" button to delete it from the Less Than Match.
Match Value Format - The value for the Less Than Match type can be entered in Decimal, HEX,
Binary or ASCII format. The value will be limited to the maximum value allowed for the selected
signal, which will depend on the number of bit positions assigned to the signal. The formats for
entering each value type is described below.
Duration - Enter the match duration by selecting the units then typing the value. The Maximum
Duration period for Model DV3400 is 5 ms. The Trigger condition will not be considered as "True"
until the entered value for the selected signal has been valid for the indicated duration period.
Identifier - Change the identifier to something meaningful for the match condition if you do not
want the default "Match #" to be displayed in the formula and the graphical configuration area of
the trigger screen.
Decimal: Enter any valid decimal value. An Invalid value will be cleared to 0 or assumed to be
HEX.
HEX: Start the value with an "x" or a "0x". If too many digits are entered, the most significant
nibble (leftmost) will be discarded. Any non-HEX digit will be ignored.
Binary: Enter "b" followed by "1's" and "0's". If too many digits are typed, the MSb (leftmost bit)
will be discarded. Any numerical value larger than a "1" or non-numerical character will be
considered a "1".
ASCII: Start the value with a single or double quote.
Configuration