5492GPIB B&K Precision, 5492GPIB Datasheet - Page 24

no-image

5492GPIB

Manufacturer Part Number
5492GPIB
Description
DMM BENCH 5 1/2DGT WITH GPIB CRD
Manufacturer
B&K Precision
Type
Digital (DMM)r
Datasheets

Specifications of 5492GPIB

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
120000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Accuracy
+/- 0.012 %
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
!
!
3-7 Selecting Measurement Range (
Ranging operations are performed by using the
(see Figure 3-3).
Measurement ranges can be selected automatically by the meter in
“Auto-ranging” or manually operated by the user.
! Auto-Ranging
! Manual Range
! Selecting A Measurement Range
Press
auto-range mode, the AUTO annunciator is lit.
In auto-range, the meter selects the next higher range automatically when a
reading is greater than full scale. If no higher range is available, ‘OL’
(overload) will be displayed on primary or secondary display. Likewise,
the meter will automatically selects a lower range when a reading is less
than approximately 9.5% of the full scale.
Press
become the selected range when user enters the manual range mode. In
manual range, the meter remains in the selected range regardless of input.
Press
either on readings shown in the primary display or secondary display.
To manually select a range,
In manual range mode,
, and
to toggle in (and out) of the manual ranging mode, or
, and
to toggle in and out of manual ranging. The range user is in will
to toggle back to auto-ranging. Manual range can be performed
to toggle in and out of manual ranging. When meter is in
to select higher range or lower range to the desired one.
24
,
,
, and
, and
buttons
)