5492GPIB B&K Precision, 5492GPIB Datasheet - Page 20

no-image

5492GPIB

Manufacturer Part Number
5492GPIB
Description
DMM BENCH 5 1/2DGT WITH GPIB CRD
Manufacturer
B&K Precision
Type
Digital (DMM)r
Datasheets

Specifications of 5492GPIB

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
120000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Accuracy
+/- 0.012 %
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
3-4 Input Terminals
All functions
(1)
(2)
Function
The input terminals, shown in Figure 3-2 are located on the left side of the front
panel. The meter is protected against overloads up to the limits shown in Table
3-1. Exceeding these limits poses a hazard to both the meter and operator.
12A, Hz
Up to 1200mA for 5492; up to 120mA only for 5491
Vac, Hz
10A dc or ac rms continuous, and 12A dc or ac rms for 30 seconds maximum
mA, Hz
Vdc
(2W)
(4W)
V
V
mA to Lo
12A to Lo
V
V
Sense Hi to Sense Lo
Any terminal to earth
Hz (Hi) to Lo
Input Terminal
Hz (Hi) to Lo
Hz to Lo
Hz to Lo
Table 3-1 Input Protection Limits
Figure 3-2 Input Terminals
!
!
!
!
!
!
Ohm (4W) Sense-High Terminal
Input-High Terminal for Volts, Ohms, Hz,
Diode/Continuity Measurements
12mA~1200mA Range DC/AC Current Input and
Ohm (4W) Sense-Low Terminal
Input-Low and Common Terminal (COM)
12A Range Current Input Terminal for 12A Range
DC/AC Current Measurement
12mA/120mA/1200mA mA Fuse & Holder (2A for 5492;
250mA for 5491)
20
1000V dc
750V ac rms, 1100V peak, 2x10
normal mode, or 1x10
1200mA
12A
500V dc or ac rms
500V dc or ac rms
250V dc or ac rms
1000V dc or peak ac
(2)
dc or ac rms
Maximum Allowable Input
(1)
dc or ac rms
6
V-Hz common mode
7
V-Hz