5492GPIB B&K Precision, 5492GPIB Datasheet - Page 114

no-image

5492GPIB

Manufacturer Part Number
5492GPIB
Description
DMM BENCH 5 1/2DGT WITH GPIB CRD
Manufacturer
B&K Precision
Type
Digital (DMM)r
Datasheets

Specifications of 5492GPIB

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
120000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Accuracy
+/- 0.012 %
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
!
Resolution, Full Scale Reading and Accuracy
Measurement method: True RMS
Maximum Crest Factor: 3.0 at full scale
Maximum input voltage: 750V rms, 1100V peak ac
Input Impedance: 1M
Response Time: Approximately 1.2 seconds when the displayed reading reaches 99.9%
frequency value of the tested input signal at the same range.
Range
1200
12k
120k
(1)
(2)
(Hz)
Frequency
Specified accuracy at input >5% of full scale
If tested frequency is greater than 1MHz, it will be displayed but no specified accuracy
is guaranteed.
Measurement
100 to 120k
Range (Hz)
10 to 12k
5 to1200
in parallel with capacitance <120pF
2x10 7 V-Hz product on any range, normal mode input
1x10 6 V-Hz product on any range, common mode input
Resolution
100m
10m
(Hz)
1
Full Scale
114
Reading
1199.99
11.9999
119.999
(1 year)
0.01 + 5
0.01 + 3
0.01 + 3
Accuracy
(1)
Input Sensitivity
(Sine wave)
50mV rms