3-12
4-Wire Resistance Measurement
The Meter incorporates two methods of making a four-wire resistance measurement. The
traditional method is to use four meter leads to connect the Meter to the resistance to be
measured. The optional 2X4 Wire test leads simplifies the four-wire measurement so you
only have to plug in two test leads to the Input HI and LO connectors of the meter.
To perform a 4-wire resistance measurement using four test leads:
To make a four-wire resistance measurement using Fluke’s 2X4 test leads:
1. Connect the test leads to the Meter’s input connectors as show in Figure 3-7.
2. Press N. 2*4 Wire is displayed.
1. Connect test leads between the Meter and the circuit under test as shown in
2. If required, press O to select 4-wire resistance measurement mode. 4 Wire is
Figure 3-6.
displayed.
The function and measurement are displayed.
Figure 3-6. 4-Wire Resistance Measurement