8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 22

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
8808A
Users Manual
1-12
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
200 A
2 mA
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
20 mA
200 mA
2 A
10 A
Range
Notes:
[1]
Notes:
[1]
Input Protection ................................................ Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method ....................................... AC-coupled True RMS
Shunt Resistance.............................................. 0.01 Ω for 2 A and 10 A ranges
AC Filter Bandwidth:
Maximum Crest Factor ..................................... 3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) ...... Crest Factor 1-2, 0.05 % of full scale
Range
Range
Slow ................................................................ 20 Hz – 100 kHz
Accuracy given as ± (% of reading + % of range)
Accuracy given as ± (% of reading + % of range)
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
45 Hz - 2 kHz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
20 Hz - 45Hz
(5-1/2 Digits)
19.9999 mA
199.999 mA
Frequency
Full-Scale
1.99999 A
10.0000 A
0.015 + 0.005
23 °C ± 5 °C
0.02 + 0.005
0.02 + 0.005
0.03 + 0.02
0.05 + 0.02
0.18 + 0.01
90 days
100 A
Accuracy
0.1 A
10 A
Slow
1 A
23 °C ± 5 °C
0.25 + 0.05
0.25 + 0.05
0.25 + 0.05
Input Characteristics
0.8 + 0.05
0.35 + 0.1
90 days
1 + 0.05
1 + 0.05
1 + 0.1
1 Ω for 20 mA and 200 mA
Crest Factor 2-3, 0.2 % of full scale
[1]
Accuracy
Accuracy
23 °C ± 5 °C
0.03 + 0.005
0.02 + 0.005
0.03 + 0.008
0.04 + 0.02
0.08 + 0.02
0.2 + 0.01
Resolution
1 year
Accuracy
Medium
100 A
10 A
1 mA
1 A
[1]
23 °C ± 5 °C
1.25 + 0.06
1.25 + 0.06
1.25 + 0.12
0.3 + 0.06
0.3 + 0.06
0.3 + 0.06
0.5 + 0.12
1 + 0. 06
1 year
100 A
10 A
1 mA
Fast
1 A
Temperature Coefficient/°C
Outside 18 – 28 °C
0.003 + 0.001
0.002 + 0.001
0.005 + 0.001
0.005 + 0.001
0.008 + 0.001
0.008 + 0.001
Outside 18 – 28 °C
Burden Voltage
Coefficient/°C
Temperature
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
<0.05 V
<0.5 V
<0.1 V
<0.5 V