2306-VS KEITHLEY, 2306-VS Datasheet - Page 5

58T8427

2306-VS

Manufacturer Part Number
2306-VS
Description
58T8427
Manufacturer
KEITHLEY
Datasheet

Specifications of 2306-VS

No. Of Outputs
2
Output Voltage
15V
Output Voltage 2
15V
Output Current
5A
Output Current 2
5A
Load Regulation
0.01%
Length
411mm
Width
213mm
Height
89mm
Weight
3.9kg
Power Supply Output Type
Variable
Input Voltage
100VAC To 120VAC / 220VAC To 240VAC
Power Rating
50W
2306-VS
Output #2 (Charger)
DC VOLTAGE OUTPUT (2 Years, 23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 10mV).
PROGRAMMING RESOLUTION: 10mV.
READBACK ACCURACY
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: ±(0.01% + 2mV).
LINE REGULATION: ±0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
TRANSIENT RESPONSE:
REMOTE SENSE: 1V max. drop in each lead. Add 2mV to the voltage load regulation specifica-
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
CONTINUOUS AVERAGE SINK CURRENT:
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
READBACK RESOLUTION: 5A Range: 100µA. 5mA Range: 0.1µA.
LOAD REGULATION: ±(0.01% + 1mA).
LINE REGULATION: ±0.5mA.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
Transient Recovery Time
Transient Voltage Drop
tion for each 1V change in the negative output lead due to load current change. Remote sense
required. Integrity of connection continually monitored. If compromised, output will turn off
automatically once settable window (±0 to ±8 volts) around normal voltage exceeded.
Channel #1 (Battery) OFF: I = 50W/(V
Channel #1 (Battery) ON: I = (50W – Power consumed by channel #1)/(V
Peak currents can be a maximum of 5A provided the average current is within the above limits.
Channel #1 (Battery) OFF:
Channel #1 (Battery) ON:
5A max.
The power consumed by channel #1 is calculated as:
Channel #1 sourcing current: Power consumed = (V
Channel #1 sinking current: Power consumed = 5 × (sink current)
0–5V: 3A max.
5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Available current = (50W – Power consumed by channel #1)/5; 3A max. (0–5V).
Derate 0.2A per volt above 5V.
1.888.KEITHLEY
w w w.keithley.com
2
4
: ±(0.01% + 50µA).
: ±(0.01% + 0.5mV).
1, 8, 9
1, 8, 9
: 31ms, typical.
: 31ms, typical.
1
1
: ±(0.05% + 3mV).
: 5A Range: ±(0.2% + 200µA). 5mA Range: ±(0.2% + 1µA).
13
<120 mV
<50 µs
High Bandwidth
3
3
set
or <80 µs
or <150 mV
channel 2 + 6V); 5A max.
(U.S. only)
7
7
, in 0.01PLC steps.
, in 0.01PLC steps.
4
5
4
.
set
channel 1 + 6V) × (current supplied)
<160 mV
<60 µs
Low Bandwidth
Dual-Channel Battery/Charger
Simulator with External Triggering
3
3
or <100 µs
or <200 mV
set
channel 2 + 6V);
4
4
A
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL: 5mA to 5A, in 5mA steps.
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
PULSE CURRENT MEASUREMENT ACCURACY
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs to 833ms, in 33µs steps.
CONVERSION RATE: 2040/second at 33.3µs meas. aper., typical.
INTERNAL TRIGGER DELAY: 15µs with 33µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
MEASUREMENT TIME
DIGITAL VOLTMETER INPUT (2 Years, 23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2MW typical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW: –5V, +30V.
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #2’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
VOLTAGE SETTLING TIMES (typical)
NOTE: Times are under no load condition and settling times defined at ±2% of step size.
G R E A T E R
500 µs – <1 PLC
100 µs – 200 µs
200 µs – 500 µs
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps.
Average Readings: 1 to 100.
0 V < Voltage step < 15 V
Decreasing Voltage
Increasing Voltage
Aperture
<100 µs
1 PLC
>1 PLC
Voltage step ≤ 7 V
Voltage step > 7 V
12
1, 8, 9
M E A S U R E
: 31ms, typical.
1
6
: ±(0.05% + 3mV).
: 850ms (840ms) to 60 seconds in 1ms steps.
±(% reading + offset + rms noise
10–90% Rise Time
10–90% Fall Time
0.2% + 900 µA +
0.2% + 900 µA +
0.2% + 900 µA +
0.2% + 600 µA +
0.2% + 400 µA +
0.2% + 400 µA + 100 µA
O F
10 µs to 1.2 ms
5 µs to 40 µs
10 µs
C O N F I D E N C E
Accuracy
11
7
, in 0.01PLC steps.
(2 Years, 23°C ±5°C):
2 mA
1.5 mA
1 mA
0.8 mA
0 mA
Settling Time
100 µs to 1.5 ms
Settling Time
50 µs to 200 µs
100 µs
10
)

Related parts for 2306-VS