2302-PJ KEITHLEY, 2302-PJ Datasheet

58T8418

2302-PJ

Manufacturer Part Number
2302-PJ
Description
58T8418
Manufacturer
KEITHLEY
Datasheet

Specifications of 2302-PJ

No. Of Outputs
1
Output Voltage
15V
Output Current
5A
Load Regulation
0.01%
Length
411mm
Width
213mm
Height
89mm
Weight
3.2kg
Power Supply Output Type
Variable
Input Voltage
100VAC To 120VAC / 220VAC To 240VAC
Power Rating
60W
2302-PJ Battery Simulator
DC VOLTAGE OUTPUT (2 YEARS, 23°C ± 5°C)
OUTPUT VOLTAGE: 0 to +15VDC.
OUTPUT ACCURACY: ±(0.05% + 3mV).
PROGRAMMING RESOLUTION: 1mV.
READBACK ACCURACY
READBACK RESOLUTION: 1mV.
OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy.
LOAD REGULATION: 0.01% + 2mV.
LINE REGULATION: 0.5mV.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
TRANSIENT RESPONSE:
REMOTE SENSE 1V max. drop in each lead. Add 2mV to the voltage load regulation
VARIABLE OUTPUT IMPEDANCE
RANGE: 0 to 1.00 in 0.01 steps. Value can be changed with output on.
DC CURRENT (2 Years, 23°C ± 5°C)
CONTINUOUS AVERAGE OUTPUT CURRENT:
CONTINUOUS AVERAGE SINK CURRENT:
SOURCE COMPLIANCE ACCURACY: ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION: 1.25mA.
READBACK ACCURACY
READBACK RESOLUTION: 5A Range:
LOAD REGULATION: 0.01% + 1mA.
LINE REGULATION: 0.5mA.
STABILITY
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
Transient Recovery Time
Transient Voltage Drop
specification for each 1V change in the negative output lead due to load current
change. Remote sense required. Integrity of connection continually monitored. If
compromised, output will turn off automatically once settable window (±0 to ±8
volts) around normal voltage exceeded.
0–4V: 5A max.
>4V: I
Peak currents can be a maximum of 5A provided the average current is within
the above limits.
0–5V: 3A max.
5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
MAX
2
4
: 0.01% + 0.5mV.
: 0.01% + 50µA.
= 60W / (V
1, 8, 9
1, 8, 9
: 31ms, typical.
: 31ms, typical.
1
SET
1
: ±(0.05% + 3mV).
:
+ 6) (Not intended to be operated in parallel).
13
5A Range:
500mA Range: ±(0.2% + 20µA).
500mA Range: 10µA.
<75mV
<40µs
High Bandwidth
3
3
or <60µs
or <100mV
±(0.2% + 200µA).
100µA.
7
7
, in 0.01PLC steps.
4
, in 0.01PLC steps.
4
5
.
<250mV
<80µs
Low Bandwidth
3
3
or <100µs
or <400mV
4
4
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL:
TRIGGER DELAY: 0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY: 15µs.
HIGH/LOW/AVERAGE MODE:
PULSE CURRENT MEASUREMENT ACCURACY
BURST MODE CURRENT MEASUREMENT
MEASUREMENT APERTURE: 33.3µs.
CONVERSION RATE: 3650/second, typical.
INTERNAL TRIGGER DELAY: 15µs.
NUMBER OF SAMPLES: 1 to 5000.
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE: 4800 bytes/s, typical.
LONG INTEGRATION MODE CURRENT MEASUREMENT
Available on both 5A and 500mA current ranges.
MEASUREMENT TIME
DIGITAL VOLTMETER INPUT (2 Years, 23°C ± 5°C)
INPUT VOLTAGE RANGE: –5 to +30VDC.
INPUT IMPEDANCE: 2M typical.
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW:
READING ACCURACY
READING RESOLUTION: 1mV.
CONNECTOR: HI and LO input pair part of Output #1’s terminal block.
MEASUREMENT TIME CHOICES: 0.01 to 10 PLC
AVERAGE READINGS: 1 to 10.
READING TIME
5A Current Range
500mA Current Range
500 µs – <1 PLC
–5V, +30V.
100 µs – 200 µs
200 µs – 500 µs
Measurement Aperture Settings: 33.3µs to 833ms, in 33.3µs steps.
Average Readings: 1 to 100.
APERTURE
5A Range:
1A Range:
100mA Range: 0.1mA to 100mA, in 100µA steps.
500mA Range: 0.5mA to 500mA, in 0.5mA steps.
100mA Range: 0.1mA to 100mA, in 100µA steps.
10mA Range:
<100 µs
1 PLC
>1 PLC
12
1, 8, 9
: 31ms, typical.
5mA to 5A, in 5mA steps.
1mA to 1A, in 1mA steps.
100µA to 10mA, in 100µA steps.
1
6
: ±(0.05% + 3mV).
0.2% + 900 µA + 2 mA
0.2% + 900 µA + 1.5 mA
0.2% + 900 µA + 1 mA
0.2% + 600 µA + 0.8 mA
0.2% + 400 µA + 0 mA
0.2% + 400 µA + 100 µA
: 850ms (840ms) to 60 seconds in 1ms steps.
5A RANGE
ACCURACY ±(% reading + offset + rms noise
11
7
, in 0.01PLC steps.
(2 Years, 23°C ±5°C):
0.2% + 90 µA + 2 mA
0.2% + 90 µA + 1.5 mA
0.2% + 90 µA + 1 mA
0.2% + 60 µA + 0.8 mA
0.2% + 40 µA + 0 mA
0.2% + 40 µA + 100 µA
500mA RANGE
HW 6/27/02
10
)
Rev. A

Related parts for 2302-PJ

2302-PJ Summary of contents

Page 1

... Battery Simulator DC VOLTAGE OUTPUT (2 YEARS, 23°C ± 5°C) OUTPUT VOLTAGE +15VDC. OUTPUT ACCURACY: ±(0.05% + 3mV). PROGRAMMING RESOLUTION: 1mV. READBACK ACCURACY 1 : ±(0.05% + 3mV). READBACK RESOLUTION: 1mV. OUTPUT VOLTAGE SETTLING TIME: 5ms to within stated accuracy. LOAD REGULATION: 0.01% + 2mV. LINE REGULATION: 0.5mV. ...

Page 2

... Battery Simulator GENERAL ISOLATION (low–earth): 22VDC max. Do not exceed 60VDC between any two ter- minals of either connector. PROGRAMMING: IEEE-488.2 (SCPI). USER-DEFINABLE POWER-UP STATES: 4. REAR PANEL CONNECTORS: 8-position quick disconnect terminal block for out- put (4), sense (2), and DVM (2). TEMPERATURE COEFFICIENT (outside 23°C ±5°C): Derate accuracy specification by (0.1 specification)/° ...

Related keywords