Datasheet

Specifications of AD5933YRSZ

Resolution (bits)
12 b
Master Fclk
16.776MHz
Voltage - Supply
2.7 V ~ 5.5 V
Operating Temperature
-40°C ~ 125°C
Mounting Type
Surface Mount
Package / Case
16-SSOP
Supply Voltage Range
2.7V To 5.5V
Operating Temperature Range
-40°C To +125°C
Digital Ic Case Style
SSOP
No. Of Pins
16
Frequency Max
0.1MHz
Termination Type
SMD
Pin Count
16
Screening Level
Automotive
Package Type
SSOP
Filter Terminals
SMD
Rohs Compliant
Yes
Communication Function
Network Analyzer
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
For Use With
EVAL-AD5933EBZ - BOARD EVALUATION FOR AD5933
Tuning Word Width (bits)
-
Lead Free Status / Rohs Status
Compliant
Other names
AD5933BRSZ
Q2204656A

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
AD5933YRSZ
Manufacturer:
ADI
Quantity:
5 000
Part Number:
AD5933YRSZ
Manufacturer:
Fujitsu
Quantity:
500
Part Number:
AD5933YRSZ
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Table 11. Control Register Map (D11, D8 to D0)
Bits
D11
D8
D7
D6
D5
D4
D3
D2
D1
D0
Control Register Decode
Initialize with Start Frequency
This command enables the DDS to output the programmed
start frequency for an indefinite time. It is used to excite the
unknown impedance initially. When the output unknown
impedance has settled after a time determined by the user, the
user must initiate a start frequency sweep command to begin
the frequency sweep.
Start Frequency Sweep
In this mode the ADC starts measuring after the programmed
number of settling time cycles has elapsed. The user has the
ability to program an integer number of output frequency cycles
(settling time cycles) to Register Address 0x8A and Register
Address 0x8B before the commencement of the measurement
at each frequency point (see Figure 34).
Increment Frequency
The increment frequency command is used to step to the next
frequency point in the sweep. This usually happens after data
from the previous step has been transferred and verified by the
DSP. When the AD5933 receives this command, it waits for the
programmed number of settling time cycles before beginning
the ADC conversion process.
Repeat Frequency
The AD5933 has the facility to repeat the current frequency
point measurement by issuing a repeat frequency command to
the control register . This has the benefit of allowing the user to
average successive readings.
Measure Temperature
The measure temperature command initiates a temperature
reading from the part. The part does not need to be in power-
up mode to perform a temperature reading. The block powers
itself up, takes the reading, and then powers down again. The
temperature reading is stored in a 14-bit, twos complement
format at Register Address 0x92 and Register Address 0x93.
Description
No operation
PGA gain; 0 = ×5, 1 = ×1
Reserved; set to 0
Reserved; set to 0
Reserved; set to 0
Reset
External system clock; set to 1
Internal system clock; set to 0
Reserved; set to 0
Reserved; set to 0
Reserved; set to 0
Rev. C | Page 25 of 44
Power-Down Mode
The default state on power-up of the AD5933 is power-down
mode. The control register contains the code 1010,0000,0000,0000
(0xA000). In this mode, both the VOUT and VIN pins are
connected internally to GND.
Standby Mode
This mode powers up the part for general operation; in standby
mode the VIN and VOUT pins are internally connected to ground.
Output Voltage Range
The output voltage range allows the user to program the
excitation voltage range at VOUT.
PGA Gain
The PGA gain allows the user to amplify the response signal
into the ADC by a multiplication factor of ×5 or ×1.
Reset
A reset command allows the user to interrupt a sweep. The start
frequency, number of increments, and frequency increment
register contents are not overwritten. An initialize with start
frequency command is required to restart the frequency sweep
command sequence.
START FREQUENCY REGISTER (REGISTER
ADDRESS 0x82, REGISTER ADDRESS 0x83,
REGISTER ADDRESS 0x84)
The default value of the start frequency register upon reset is
as follows: D23 to D0 are not reset on power-up. After a reset
command, the contents of this register are not reset.
The start frequency register contains the 24-bit digital represen-
tation of the frequency from where the subsequent frequency
sweep is initiated. For example, if the user requires the sweep to
start from frequency 30 kHz (using a 16.0 MHz clock), then the
user programs the value of 0x0F to Register Address 0x82, the
value of 0x5C to Register Address 0x83, and the value of 0x28 to
Register Address 0x84. This ensures the output frequency starts
at 30 kHz.
The code to be programmed to the start frequency register is
Start
Frequency
Code
=
16
30
MHz
kHz
4
×
2
27
0
0 x
AD5933
5 F
C
28

Related parts for AD5933YRSZ