U3606A AGILENT TECHNOLOGIES, U3606A Datasheet - Page 9

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

U3606A

Manufacturer Part Number
U3606A
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheets

Specifications of U3606A

Dmm Type
Bench
Voltage Measuring Range Dc
100mV To 1000V
Voltage Measuring Range Ac
100mV To 750V
Current Measuring Range Dc
10mA To 3A
Current Measuring Range Ac
10mA To 3A
Lead Free Status / RoHS Status
na
Digital multimeter specifications (continued)
Supplementary specifications
Diode
Measurement method
Open circuit voltage
Audible tone
Input protection
Capacitance
Measurement method
Maximum voltage at full scale
Input protection
Response time
Charge and discharge voltage
Frequency
Measurement method
Signal level
Input protection
Maximum display counts (excluding frequency)
5½ digits
4½ digits
Noise rejection
Common mode rejection ratio (CMRR)
for 1 kΩ unbalanced in LO lead
Normal mode rejection ratio (NMRR)
0.83 mA ± 0.2% constant current source
< +5.0 V
• Continuous beeping when level is below +50 mV DC
• Single tone for normal forward-biased diode or semiconductor junction where
1000 V
Computed from constant current source charge time, typical 0.2 V to 1.4 V
• For 1 nF to 10 µF range: < 1.5 V
• For 100 µF to 10000 µF: 0.33 V
1000 V
Approximately 1 s for 100 µF and ranges below 100 µF
5 V
Reciprocal counting technique
0.2 V to 1.4 V
• Voltage path: 1000 V
• Current path: Protected with 3.15 A/500 V, FF fuse
120,000
12,000
• DC: 140 dB
• AC: 70 dB
• 60 Hz ± 0.1%:
• 50 Hz ± 0.1%:
0.3 V ≤ reading ≤ 0.8 V
5½ digits: 65 dB, 4½ digits: 0 dB
5½ digits: 55 dB, 4½ digits: 0 dB
pp
(approximately from +3 V to –2 V)
rms
rms
dc
on all ranges, < 0.3 A short circuit
on all ranges, < 0.3 A short circuit
rms
9
on all ranges, < 0.3 A short circuit
ac
signal level