U3606A AGILENT TECHNOLOGIES, U3606A Datasheet

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

U3606A

Manufacturer Part Number
U3606A
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheets

Specifications of U3606A

Dmm Type
Bench
Voltage Measuring Range Dc
100mV To 1000V
Voltage Measuring Range Ac
100mV To 750V
Current Measuring Range Dc
10mA To 3A
Current Measuring Range Ac
10mA To 3A
Lead Free Status / RoHS Status
na
Features
• Two independent instruments
• Nine basic measurements
• Measurement speed: up to
• Low error rate: up to 0.025%
• 30-W dual-range power supply
• Excellent load regulation: up to
• OVP and OCP load protection
• Auto scan and ramp, and
• USB-TMC488.2 and GPIB
• Kensington lock slot security
in one box
as well as 4-wire milliohm
measurement and eight math
functions
37 readings/s
DCV accuracy
with remote sensing
0.01%+3 mV
built-in 4.8 kHz square-wave
generator
connectivity
Agilent U3606A
Multimeter|DC Power Supply
Data Sheet
The real two instruments in one box
Pick both
Digital multimeter (DMM) and power
supply? Simultaneous and inde-
pendent? Small and cost-effective?
Get all this and more in the Agilent
U3606A multimeter|DC power supply.
This convenient new hybrid combines
a 5½-digit DMM and 30-W dual-range
supply in a single unit. Operating
simultaneously and independently,
the instruments provide efficient,
affordable testing while saving space
on the bench or in a rack.
The 5.5-digit DMM
The 5.5-digit DMM includes nine
essential multimeter capabilities as
well as 4-wire milliohm measurement
and eight built-in math functions. The
DMM also provides a fast measure-
ment speed of up to 37 readings/s,
and a low error rate of up to 0.025%
DCV accuracy.
Physical security and seamless
system integration
Instruments may be at risk of theft or
misplacement when left unattended
on the bench. With the hybrid multi-
meter’s rear Kensington lock slot, you
can secure your instrument and be
assured that it is where you expect it
to be for your continued testing the
next day. The rackmountable U3606A
also enables seamless integration
into your system via popular GPIB
and USB-TMC488.2 interfaces,
programmable with standard SCPI
commands.

U3606A Summary of contents

Page 1

... Kensington lock slot, you can secure your instrument and be assured that it is where you expect for your continued testing the next day. The rackmountable U3606A also enables seamless integration into your system via popular GPIB and USB-TMC488.2 interfaces, programmable with standard SCPI commands ...

Page 2

... Sweep functions Sweep functions in the U3606A are auto ramp and scan outputs for low- speed multilevel DC bias tests, such as margin tests, power cycling tests and relay control. Both functions are ...

Page 3

... Take a closer look Source terminals Multimeter terminals Figure 1. Front panel of the U3606A. AC line fuse AC power connector Figure 2. Rear panel of the U3606A. VFD display Multimeter Autorange operation keys and manual range keys Kensington GPIB (IEEE-488) Ventilation fan lock slot connector Chassis Rear output terminals ...

Page 4

... Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset current im- mediately after a high-current measurement, ensure that the U3606A has cooled down. 4. Specifications stated are for 2-wire resistance measurements using Null math operation. Without Null, add a 0.2 Ω error. To eliminate noise interfer- ence which may be induced by the test leads, a shielded test cable is recommended for resistances above 100 kΩ ...

Page 5

Digital multimeter specifications (continued) Low-resistance specifications Range Test current 100 mΩ 1.0000 A 1000 mΩ 0.1000 A 10 Ω 0.1000 A 1. Four-wire measurement method is used. Test current is sent from the FORCE terminals and resistance is measured at ...

Page 6

Digital multimeter specifications (continued) Frequency specifications Function Input range Frequency Voltage: 100 mV to 750 V 1 Current Minimum frequency sensitivity for voltage measurement (RMS sine wave) 2 Input range 100 kHz ...

Page 7

Digital multimeter specifications (continued) Measurement speeds (typical) Function Rate DC voltage Slow (5½ digits) (10 V) Fast (4½ digits) DC current Slow (5½ digits Fast (4½ digits) AC voltage Slow (5½ digits) ( kHz) Fast ...

Page 8

Digital multimeter specifications (continued) Supplementary specifications AC voltage Measurement method Maximum input voltage Input impedance Input protection Crest factor Peak input Response time Overload ranging AC current Measurement method Maximum input current Burden voltage and shunt resistance Input protection Crest ...

Page 9

Digital multimeter specifications (continued) Supplementary specifications Diode Measurement method Open circuit voltage Audible tone Input protection Capacitance Measurement method Maximum voltage at full scale Input protection Response time Charge and discharge voltage Frequency Measurement method Signal level Input protection Maximum ...

Page 10

... Overcurrent protection (for CV mode) 1. Contacts and leads resistance may contribute an additional error of 6 mV/A (typical). 2. Average time for the detection of OVP or OCP condition. 3. Refer to U3606A User’s and Service Guide (U3606-90013); CV noise effect measurement on page 166. • Range S1 • Range S2 • 0.05 • ...

Page 11

Power supply specifications (continued) Sweep specifications Function Range Scan Constant voltage S1 (30 V V/3 A) Constant current S1 (30 V V/3 A) Ramp Constant voltage S1 (30 V V/3 A) ...

Page 12

General characteristics Power supply Power consumption Current Input Fuse Display Operating environment Storage compliance Safety compliance EMC compliance Shock and vibration Remote interface Programming language Measurement category Dimensions (H × W × D) Weight Warranty Calibration cycle Warm-up time • ...

Page 13

... Kelvin Clip Set (for DMM function) Warranty options • R-51B-001-3C Extended warranty from one year to three years • R-51B-001-5C Extended warranty from one year to five years U3606A-1CM Rack Mount Kit 34330A Current Shunt (30 A) (for DMM function) E3600A-100 Test Lead Kit (for DC power supply function) 13 ...

Page 14

... Spain 34 (91) 631 3300 Sweden 0200- United Kingdom 44 (0) 118 9276201 For other unlisted Countries: www.agilent.com/find/contactus Revised: October 14, 2010 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2011 Printed in USA, April 4, 2011 5990-3971EN ...