U3606A AGILENT TECHNOLOGIES, U3606A Datasheet - Page 6

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

U3606A

Manufacturer Part Number
U3606A
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheets

Specifications of U3606A

Dmm Type
Bench
Voltage Measuring Range Dc
100mV To 1000V
Voltage Measuring Range Ac
100mV To 750V
Current Measuring Range Dc
10mA To 3A
Current Measuring Range Ac
10mA To 3A
Lead Free Status / RoHS Status
na
Digital multimeter specifications (continued)
Frequency specifications
1. For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V input signal.
2. Maximum input for the specified accuracy = 10 x range or 1000 V
All frequency counters are susceptible to errors when measuring low-voltage, low-frequency signals. Shielding inputs from
external noise pickup is critical for minimizing measurement errors.
Duty cycle and pulse width specifications
1. The range is from {10 µs × frequency × 100%} ~ {[1 – (10 µs × frequency)] × 100%}. For example, a 1 kHz signal can be measured from
2. The positive or negative pulse width must be greater than 10 µs. The range of the pulse width is determined by the frequency of the signal.
Function
Frequency
Input range
100 mV
1.0 V
10 V
100 V
750 V
Input range
10 mA
100 mA
1 A
3 A
Function
Duty cycle
Pulse width
1% ~ 99%.
2
Input range
Voltage:
100 mV to 750 V
Current:
10 mA to 3 A
Minimum frequency sensitivity for voltage measurement (RMS sine wave)
20 Hz to 100 kHz
50 mV
100 mV
1 V
10 V
100 V
Minimum frequency sensitivity for current measurement (RMS sine wave)
20 Hz to 100 kHz
1 mA
10 mA
100 mA
300 mA
Range
100.000%
199.999 ms
1999.99 ms
1
1
2
2
Frequency range:
minimum input
frequency = 1 Hz
< 2 Hz
< 20 Hz
20 Hz to 100 kHz
100 kHz to 300 kHz
< 2 Hz
< 20 Hz
20 Hz to 10 kHz
dc
Resolution
0.001%
0.001 ms
0.01 ms
.
100 kHz to 300 kHz
50 mV
120 mV
1.2 V
12 V
6
Accuracy
±(% of reading + % of range)
1 year (23 ºC ± 5 ºC)
0.18 + 0.003
0.04 + 0.003
0.02 + 0.003
0.02 + 0.003
0.18 + 0.003
0.04 + 0.003
0.02 + 0.003
Accuracy at full scale
0.3% + 0.2% per kHz
Duty cycle/frequency
Duty cycle/frequency
300 kHz to 1 MHz
0.5 V
0.5 V
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
0.005
0.005
0.005
0.005
0.005
0.005
0.005