FLUKE-289/FVF Fluke Electronics, FLUKE-289/FVF Datasheet - Page 78

COMBO KIT DMM & ACCESSORY

FLUKE-289/FVF

Manufacturer Part Number
FLUKE-289/FVF
Description
COMBO KIT DMM & ACCESSORY
Manufacturer
Fluke Electronics
Series
FlukeView® Forms, 28xr
Type
Digital (DMM)r
Datasheet

Specifications of FLUKE-289/FVF

Includes
Battery, Case, Clips, Software, Temperature Probe, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
50000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB
Features
Data Logging, Hold, Memory, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
289/FVF
614-1063
Low pass filter
287/289
Users Manual
AC Voltage Specifications
68
AC mV
AC V
dBV
L
[1]
[2]
[3]
[4]
[5]
[6]
[7]
See Detailed Specifications introduction for additional information.
Function
[4]
Below 5 % of range, add 20 counts.
Specification increases linearly from -2 % at 200 Hz to -6 % at 440 Hz. Range is limited to 440 Hz.
dBm (600 ) is specified by adding +2.2 dB to the dBV range values.
289 only.
500 mV range specification between 64.000 kHz and 67.000 kHz is +(0.0 % to -5 % of reading); 50 mV range specification between 64.000 kHz and
67.000 kHz is +(0.0 % to -6 % of reading 40 counts). Valid from -20 C to <18 C and >28 C to +55 C.
Add 2.5 % above 65 kHz.
Range is limited to 440 Hz.
[5]
[4]
50 mV
500 mV
5 V
50 V
500 V
1000 V
-70 to -62 dB
-62 to -52 dB
-52 to -6 dB
-6 to +34 dB
34 to 60 dB
1000 V
Range
[1]
[1]
[1]
[1]
[3]
[3]
[3]
[3]
[3]
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
0.01 dB
0.01 dB
0.01 dB
0.01 dB
0.01 dB
0.1 V
Resolution
20 to 45 Hz
1.5 % + 60
1.5 % + 60
1.5 % + 60
1.5 % + 60
1.5 % + 60
1.5 % + 60
2 % + 80
2 % + 80
1.5 dB
0.2 dB
0.2 dB
0.2 dB
3 dB
45 to 65 Hz
0.3 % + 25
0.3 % + 25
0.3 % + 25
0.3 % + 25
0.3 % + 25
0.3 % + 25
2 % + 40
2 % + 40
1.5 dB
1.0 dB
0.1 dB
0.1 dB
0.1 dB
65 Hz to 10 kHz
Accuracy
0.4 % + 25
2 % + 40
0.4 % + 25
0.6 % + 25
0.4 % + 25
0.4 % + 25
0.4 % + 25
2 % +10
-6 % -60 [2]
0.1 dB
0.1 dB
0.1 dB
2 dB
1 dB
[7]
10 to 20 kHz
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
0.7 % + 40
0.7 % + 40
1.5 % + 40
0.7 % + 40
0.2 dB
0.2 dB
2 dB
1 dB
20 to 100 kHz
3.5 % + 40
3.5 % + 40
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
Not Spec’d
3.5 % + 40
3.5 % + 40
0.8 dB
0.8 dB
3 dB
2 dB
[6]