71m6515h Maxim Integrated Products, Inc., 71m6515h Datasheet - Page 26

no-image

71m6515h

Manufacturer Part Number
71m6515h
Description
71m6515h Energy Meter Ic
Manufacturer
Maxim Integrated Products, Inc.
Datasheet

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
71m6515h-IGT
Manufacturer:
TERIDIAN
Quantity:
317
Part Number:
71m6515h-IGT
Manufacturer:
TERIDIA
Quantity:
20 000
Part Number:
71m6515h-IGT/F
Manufacturer:
Maxim Integrated
Quantity:
10 000
Part Number:
71m6515h-IGTR/F
Manufacturer:
Maxim Integrated
Quantity:
10 000
Part Number:
71m6515h-IGTW
Manufacturer:
TERIDIAN
Quantity:
80
Part Number:
71m6515h-IGTW
Manufacturer:
TERIDIAN
Quantity:
20 000
Part Number:
71m6515h-IGTW/F
Manufacturer:
Maxim Integrated
Quantity:
10 000
Subtract 8 from 7:
9.
use equation 5:
10.
11.
12.
Now that we know the A
We calculate PHADJ from φ
Finally, we calculate the new calibration current gain coefficient, including compensation for a slight gain increase in the phase
calibration circuit.
Alternative Calibration Procedures
It is possible to implement a fast calibration based on only one measurement with a zero-degree load angle. Details can be
found in the TERIDIAN Application Note AN_651X_022 (Calibration Procedures).
Calibration Procedure for Rogowski Sensor
Rogowski coils generate an output signal that is the derivative of the input current. The 6515H Rogowski module implemented
in the Rogowski CE image digitally compensates for this effect and has the usual gain and phase calibration adjustments.
Additionally, calibration adjustments are provided to eliminate voltage coupling from the sensor input.
Current sensors built from Rogowski coils have relatively high output impedances that are susceptible to capacitive coupling
from the large voltages present in the meter. The most dominant coupling is usually capacitance between the primary of the
coil and the coil’s output. This coupling adds a component proportional to the derivative of voltage to the sensor output. This
effect is compensated by the voltage coupling calibration coefficients.
As with the CT procedure, the calibration procedure for Rogowski sensors uses the meter’s display to calibrate the voltage
path and the pulse outputs to perform the remaining energy calibrations. The calibration procedure must be performed to each
phase separately, making sure that the pulse generator is driven by the accumulated real energy for just that phase. In other
words, the pulse generator input should be set to WhA, WhB, or WhC, depending on the phase being calibrated. The IC has to
be configured for Rogowski mode (IMAGE=01). In preparation of the calibration, all calibration parameters are set to their
default values. VMAX and IMAX are set to reflect the system design parameters. WRATE and PULSE_SLOW, PULSE_FAST are
adjusted to obtain the desired Kh.
For details on calibrating a meter for Rogowski coil sensors, see the TERIDIAN Application Note AN_6515_036.
Page: 26 of 57
CAL
CAL
PHADJ
_
_
φ
E
E
E
I
V
S
NEW
60
60
60
NEW
=
=
2
tan
=
20
E
E
E
=
CAL
300
300
300
1
1 (
⎜ ⎜
CAL
A
tan(
=
XI
=
=
_
A
2
tan(
E
(
2
I
XV
60
E
_
0
XV
(
A
9
cos(
E
0
)
)(
V
+
, A
1
XV
φ
sin(
60
+
E
+
S
S
E
XI
, the desired phase lag:
A
1 )
0
180
φ
E
2
[
, and φ
XI
+
2
E
S
180
+
π
20
E
)
300
+
tan(
f
1 (
PHADJ
180
0
+
)
T
2
S
)
) 2
+
tan(
60
errors, we calculate the new calibration voltage gain coefficient from the previous ones:
2
1
) 2
© 2005-2008 TERIDIAN Semiconductor Corporation
tan(
tan(
9
)
⎟ ⎟
60
2
)
sin(
2 (
2
1 (
)
φ
60
+
sin(
S
φ
2
2
2
1 )
)
1 (
S
[
tan(
20
9
)
φ
)
PHADJ
S
cos(
2
1 (
)
φ
1
9
S
)
2
)
2
cos(
π
f
9
0
T
)
cos(
2
2
)
π
1 (
+
f
0
1 (
T
2
π
2
)
f
]
2
9
0
T
)
9
cos(
)
)
]
2
2
π
f
0
T
))
Energy Meter IC
DATA SHEET
71M6515H
MARCH 2008
V1.4

Related parts for 71m6515h