PTC-04-DB-CALIB Melexis Inc, PTC-04-DB-CALIB Datasheet - Page 16

no-image

PTC-04-DB-CALIB

Manufacturer Part Number
PTC-04-DB-CALIB
Description
DUAGHTER BOARD CALIBRATION
Manufacturer
Melexis Inc
Series
-r

Specifications of PTC-04-DB-CALIB

Accessory Type
Calibration Software
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
For Use With/related Products
PTC Programmer
What do we calibrate?
Sensing lines:
There are four sensing lines with independent ground connection.
The source and reference that are used to calibrate the sensing lines are PPS 1 and the external
voltmeter.
PPS1:
For each PPS, the software checks the ADC first then the DAC and then the additional channels that are
connected to that source.
PPS2:
PPS3:
PPS4:
User manual calibration software PTC04
UI_0PTC04AAMLX Rev. 1.35
Sensing line 1
Sensing line 2
Sensing line 3
Sensing line 4
Voltmeter on the PPS1
PPS1
Current meter on the PPS1
PPS 5: IDD limiter PPS1
Voltmeter on the PPS2
PPS2
Current meter on the PPS2
PPS 6: IDD limiter PPS2
Voltmeter on the PPS3
PPS3
Current meter on the PPS3
PPS7: IDD limiter PPS3
Voltmeter on the PPS4
PPS4
Current meter on the PPS4
PPS8: IDD limiter PPS4
Calibration Software PTC04
channel 12 and Channel 28 (High sensitivity) of the ADC;
channel 13 and Channel 29 (High sensitivity) of the ADC;
channel 14 and Channel 30 (High sensitivity) of the ADC;
channel 15 and Channel 31 (High sensitivity) of the ADC.
channel 0 and Channel 16 (High sensitivity) of the ADC;
channel 0 of the DAC;
channel 1 of the ADC;
channel 4 of the DAC.
channel 2 and Channel 18 (High sensitivity) of the ADC;
channel 1 of the DAC;
channel 3 of the ADC;
channel 5 of the DAC.
channel 4 and Channel 20 (High sensitivity) of the ADC;
channel 2 of the DAC;
channel 5 of the ADC;
channel 6 of the DAC.
channel 6 and Channel 22 (High sensitivity) of the ADC;
channel 3 of the DAC;
channel 7 of the ADC;
channel 7 of the DAC.
Page 16 of 17