OS532E NEWPORT ELECTRONICS, OS532E Datasheet - Page 59

INFRARED THERMOMETER, -18°C TO 538°C

OS532E

Manufacturer Part Number
OS532E
Description
INFRARED THERMOMETER, -18°C TO 538°C
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS532E

Temperature Measuring Range
-18°C To 538°C
Temperature Tester Type
Thermocouple - Type K
Accuracy %
1%
Wavelength Max
670nm
7-2
7
Troubleshooting Guide
The thermometer is
“locked up” (the
display is “frozen”).
The display is either
erratic or stays at
one reading.
You can expect to see and hear the following:
Problem
Remove and reinstall the batteries or
disconnect and reconnect the ac
adapter.
1. Clean the thermometer lens.
2. Activate the Diagnostic routine of the
a. Pull the trigger and press the
b. Press the
• You will see the model and version
• You will hear a beep, “TST” is displayed.
• Soon after, all of the segments of the
• The display will clear and a PAS (pass)
number “VER X.X” of the software for
about 1 second.
display including the backlighting will
light up for about 1 second.
or ERR (error) code may be seen on the
display.
ERR1: Infrared temp reading is >150°F
or < 23°F.
ERR2: Ambient temp >122°F or < 32°F
ERR3: Can not read from EEPROM
memory
EER4: Can not write to EEPROM memory
Refer to Section 4.2.
thermometer as follows (while looking
at room temp):
to lock the trigger.
same time.
FUNC
Solution
key and
LOCK
key at the
LOCK
key