OS532E NEWPORT ELECTRONICS, OS532E Datasheet - Page 51

INFRARED THERMOMETER, -18°C TO 538°C

OS532E

Manufacturer Part Number
OS532E
Description
INFRARED THERMOMETER, -18°C TO 538°C
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS532E

Temperature Measuring Range
-18°C To 538°C
Temperature Tester Type
Thermocouple - Type K
Accuracy %
1%
Wavelength Max
670nm
3-4
3
Laser Sighting
The line of sight of the thermometer does not coincide with that of
the Laser Sighting, as shown in Figure 3-4. The two lines of sight
become less critical when measuring distant targets. For example, at
30 feet from the target and a 3 foot diameter target size, there is a
2.7% offset error with respect to the target size. For close-up targets,
first make sure the target fills the laser circle, then point it with the
center of the beam approximately 1" below the center of the target.
A simple method to make infrared measurements is to scan the laser
beam across the target area vertically and horizontally and recall
measurements of maximum for hot and minimum for cold target
(compared to the background) to obtain the correct temperature.
Figure 3-4 Lines of Sight of the Laser Sighting and Thermometer
The Laser Sighting turns on only when used with the
thermometer. The module does not turn on by itself.