Specifications of 72-1015

Dmm Type
Bench
Voltage Measuring Range Dc
600mV To 1000V
Voltage Measuring Range Ac
600mV To 1000V
Current Measuring Range Dc
60µA To 10A
Current Measuring Range Ac
60µA To 10A
Lead Free Status / RoHS Status
na
Sleep Mode
The meter will automatically enter the sleep mode after
approximately 10 minutes of inactivity, to preserve
battery life.
The Meter can be activated by pressing the POWER
button off and then on or pressing the HOLD button, it
will display the last measurement value before it entered
sleep mode and under the HOLD mode.
Turning the rotary switch can also activate the Meter.
However, it will start from the switch selected function
and it will not display the last measurement value before
it entered sleep mode.
To disable the Sleep Mode function, press MAX MIN,
RANGE, or RS232 button while turning on the Meter,
the
36
symbol disappears.
The last value will be stored.
RS232 Button
Press RS232 button to enter or exit data output mode.
In RS232C serial port data output mode, the Hold and
Max Min mode cannot output to the computer, the
computer will only display the present measuring value.
In RS232C serial port data output mode, Sleep Mode
function will be disabled,
hFE and
cannot output to the computer.
Model 72-1015: OPERATING MANUAL
symbol disappears. +DC,