Model 72-1015: OPERATING MANUAL
Operation of Hold Mode
To avoid possibility of electric shock, do not use Hold
mode to determine if circuits are without power. The
Hold mode will not capture unstable or noisy readings.
The Hold mode is applicable to all measurement functions.
l
l
l
The POWER Button
This is a self-lock switch used to turn on or off the power
Meter. It is located at the rear of the Meter.
"I" indicates power on, while "O" indicates power off.
Press HOLD to enter Hold mode.
Press HOLD again to exit Hold mode and the Meter
displays the present measurement value.
In Hold mode,
Warning
is displayed.
The SELECT Button
Some positions of the rotary selector have more than
one function. This button is used for selecting the
second function.
33