BK2534 B&K Precision, BK2534 Datasheet - Page 69

OSCOPE 60MHZ DIGITAL STORAGE

BK2534

Manufacturer Part Number
BK2534
Description
OSCOPE 60MHZ DIGITAL STORAGE
Manufacturer
B&K Precision
Type
Bench, Digitalr
Datasheets

Specifications of BK2534

Includes
Probes, Software, USB Cable
Bandwidth
60MHz
Channels
Dual
Display Type
LCD - Color
Features
Software Controllable, USB Port, Waveform Capture
Probe Type
PR-37A x1/x10/Ref Probes (2)
Sampling Rate (per Second)
400MSa/s
Input Impedance
1M - 19pF
Rise Time (typ)
5.83nS
Voltage - Input, Max
400Vpeak
Voltage - Supply
100 ~ 240VAC
Lead Free Status / RoHS Status
Not applicable / Not applicable
 
Application Examples
137
138
Application Examples
PASS/FAIL Measurement
The oscilloscope measures and compares the input signal
with a predefined Pass/Fail mask. If the input signal is within
the mask threshold, a PASS verdict will be generated. If the
input signal lies outside the threshold, a FAIL verdict will be
generated.
Perform the following steps to make a PASS/FAIL
measurement.
 
Press the UTILITY key to display the UTILITY menu
page 1/2.
Press the More 1/2 softkey to display the UTILITY menu
page 2/2.
Press the Pass/Fail softkey to display the PASS/FAIL
menu.
Press the Enable Test softkey to turn on the PASS/FAIL
measurement.
Press the Setup Mask softkey to display the MASK
menu.
Press the X Mask softkey and then rotate the entry knob
to setup the horizontal threshold.
Press the Y Mask softkey and then rotate the entry knob
to setup the vertical threshold.
Press the Create Mask softkey to update the thresholds.
Press the
softkey to return to the PASS/FAIL menu.