N6752A AGILENT TECHNOLOGIES, N6752A Datasheet - Page 4

POWER SUPPLY, BENCH, 50V, 100W

N6752A

Manufacturer Part Number
N6752A
Description
POWER SUPPLY, BENCH, 50V, 100W
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of N6752A

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
50V
Output Current
10A
Power Rating
100W
Input Voltage
100V AC To 240V AC
Length
596.9mm
Width
432.5mm
Height
44.45mm
Lead Free Status / RoHS Status
na
Option 750, External Relay Control
and RI/DFI (RI latches)
The 662xA with Option 750
has an 8-pin port dedicated
to the control of up to four
external relays and a 4-pin
port dedicated to RI/DFI. The
N6700A mainframe comes
standard (no option required)
with a 4-pin port and the
N6700B mainframe comes
standard with an 8-pin port
that are user programmable to
one of three functions— RI/DFI,
digital I/O, or triggering. Digital
I/O and triggering are not
available in the 662xA.
External Relay Control
External relay control as
provided in the 662xA with
Option 750 is not supported
by the N6700A/B. The 4-pin
digital port on the N6700A,
when configured for digital
I/O, can support control of up
to 3 external relays, not 4 as
in the 662xA. The 8-pin digital
port on the N6700B, when
configured for digital I/O,
can support control of up
to 7 external relays. The
compatibility commands
associated with external relay
control are not implemented
and will produce an error.
If you would like to control
EXTERNAL relays through
the digital port, the Standard
Commands for Programmable
Instruments (SCPI) language
must be used. Refer to the
N6700 User’s Guide for infor-
mation on how to configure the
digital I/O port for this use. If
interested in INTERNAL relays
for the N675xA and N676xA
modules, Option 761, INTER-
NAL output disconnect relays,
may meet your needs. Also,
Option 054, high-speed test
extensions, which provides
digitized output measurements
and output list capability, is
available for these modules.
Option 054 is standard on the
N676xA modules.
RI/DFI
When the N6700A/B’s digital
I/O port is configured for
RI/DFI, the operation is the
same as in the 662xA with
Option 750. Please see the
N6700 User’s Guide for
information on how to
configure the digital I/O
port for RI/DFI.
4
Option S50, Relay Control and
RI/DFI (RI does not latch)
The 662xA with Option S50 is
the same as Option 750 shown
above, but the Remote Inhibit
(RI) does not latch.
External Relay Control
External Relay Control is the
same as described in Option 750.
RI/DFI
When the N6700A/B’s digital
I/O port is configured for RI/DFI,
the non-latching RI functionality
provided in the 662xA with
Option S50 can be enabled
on the N6700A/B by using the
“OUTPut:INHibit:MODE LIVE”
command. This functionality
has been made available in the
N6700A/B firmware. The default
mode for RI is “OUTPut:
INHibit:MODE LATChing”.

Related parts for N6752A