Engineering
Scaling to
8
Display
Units
PRESS
‘MENU’
‘MIN’
If your meter displays an offset which is a positive number, you
must enter a negative “–” offset number into Reading Offset
(“RDG OF”). An example would be if your zero input shows “20”
on your meter display, you would enter “–00020” in Reading Offset
(“RDG OF”) and if your zero input shows a negative “–20” you
would enter a positive 000020 in Reading Offset (“RDG OF”).
PRESS
‘MIN’
‘MENU’
‘RESET’ ‘RESET’
THEN PRESS
(TO CHANGE DISPLAY
IF REQUIRED) SHOWS
THEN PRESS
(TO CHANGE DISPLAY
IF REQUIRED) SHOWS
‘MAX’
8-22
UNTIL
“RDG OF”
“000000.”
UNTIL
“XXXXXX”
“STORED”
then
“IN CNF”
“RESET2”
then the
Measured
Value
COMMENTS
Press the ‘MENU’
button until the
display shows
“RDG OF”.
Press the ‘MIN’
button and the display
will show the last
offset entered.
COMMENTS
Use the ‘MIN’ button
to move to each digit
and the ‘MAX’ button
to change the flashing
digits value and enter
your zero offset number.
Press the ‘MENU’
button to store your
selection. The display
will momentarily
show “STORED” then
“IN CNF”.
Press the ‘RESET’
button two times. The
display will momen-
tarily show “RESET2”
and then will display
the currently
measured values.