P6021 Instruction Manual
6. Connect the probe to the current probe calibration fixture. Figure 7 shows
Figure 7: Connecting the probe to the calibration fixture
7. The following adjustments affect aberrations and flat-top response:
Calibration fixture
the proper probe orientation to the current probe calibration fixture.
C13, R10, and R12 of the probe (see Figure 8) and C22, C25, and R36 of the
termination (see Figure 9). To minimize aberrations and achieve the best
flat-top response, the following procedure is recommended. However, these
adjustments interact; you may have to readjust several of them to ensure
minimum aberrations..
a. Adjust R10 and R12 for flat response (see Figure 8). (Controls overshoot
b. Adjust C13 for flat response. (Controls slope in first 20 ns.)
c. Adjust R36 for flat response (see Figure 9). (Controls front corner, first
d. Adjust C25 to minimize aberrations in the 10 to 20 ns domain.
e. Adjust C22 to minimize aberrations in the 15 to 25 ns domain.
in the first 10 ns time domain.)
5 ns.)
Probe
Service Information
21