7. Connect the P6021 termination and probe to the BNC-to-dual banana
8. Connect the probe to the calibration fixture.
9. Record the DMM output as M2.
10. Calculate the % of error:
11. Record the results in the test record and compare the results on page 19
12. Set the P6021 termination sensitivity to 10 mA/mV.
13. Record the DMM output as M3.
14. Calculate the % of error:
15. Record the results in the test record and compare the results on page 19
Check Low Frequency Response. Refer to Figure 5 when making equipment
connections.
1. Connect the calibration fixture to the Leveled Sinewave Output of the
2. Connect the BNC-to-dual banana adapter to the input of the DMM.
3. Connect the P6021 termination and probe to the BNC-to-dual banana adapter
4. Set the DMM to ACV (autorange).
5. Set the P6021 termination sensitivity to 2 mA/mV.
6. Set the Leveled Sinewave Output of the calibration generator to 50 kHz and
7. Connect the probe to the calibration fixture.
8. Enable the Leveled Sinewave Output of the calibration generator.
adapter.
against the midband specification on page 7.
against the midband specification on page 7.
calibration generator.
connected to the DMM.
an amplitude of 5 V
Itest = M1
%Error =
%Error =
50 Ω
pp
.
2 × M2 − Itest
10 × M3 − Itest
Itest
Itest
× 100
× 100