I310S Fluke Electronics, I310S Datasheet
I310S
Specifications of I310S
I310S Summary of contents
Page 1
... Introduction The i310s Current Clamp (“Clamp”) has been designed for use with oscilloscopes and digital multimeters for accurate non- intrusive measurement of ac, dc and complex waveform currents. Using advanced Hall Effect technology, the Clamp can accurately measure currents up to 450 A peak over the frequency range kHz ...
Page 2
Safety Instructions Please read this section carefully. It will make you familiar with the most important safety instructions for handling your product. In this instruction sheet, a Warning identifies conditions and actions that pose hazard(s) to the user. A Caution ...
Page 3
CAT III equipment is designed to protect against the transients in the equipment in fixed equipment installations, such as distribution panels, feeders and short branch circuits, and the lighting systems in large buildings. Do not use Clamp in wet environments ...
Page 4
General Characteristics Maximum Conductor Size Output Cable and Connections Output Zero Cable Length Operating Temperature Range Storage Temperature Range (with battery removed) Operating Humidity Weight Safety Standards EN 61010-1: 2001 EN 61010-2-032: 2002 EN 61010-031: 2002 300 V rms Category ...
Page 5
... Typical Frequency Response i310s Frequency in Hz Typical Frequency Response Typical Accuracy i310s 30A range Typical Frequency Response Typical Accuracy i310s 300A range Typical Accuracy Curve 300A Range 30A Range evy01_4.eps Amps evy02.eps Amps evy03.eps ...
Page 6
... Operating Instructions Figure 1. i310s AC/DC Current Clamp To avoid injury, when using the Clamp, ensure that your fingers are behind the protective barrier as shown in Figure 1. Do not use the Clamp if any part, including the lead and connector(s), appears to be damaged malfunction of the instrument is suspected. ...
Page 7
Current Measurement 1. Switch on the Clamp to the required current range and check that the LED is lit. 2. Connect the output lead to an oscilloscope, multimeter, or other measuring equipment necessary, adjust the Clamp output voltage ...
Page 8
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, ...