DS1921Z-F5# Maxim Integrated Products, DS1921Z-F5# Datasheet - Page 17

IBUTTON THERMOCHRON F5

DS1921Z-F5#

Manufacturer Part Number
DS1921Z-F5#
Description
IBUTTON THERMOCHRON F5
Manufacturer
Maxim Integrated Products
Series
iButton®r
Datasheets

Specifications of DS1921Z-F5#

Rohs Information
IButton RoHS Compliance Plan
Memory Size
512B
Memory Type
NVSRAM (Non-Volatile SRAM)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Figure 7. Histogram Bin and Temperature Cross-Reference
For the temperature histogram, the DS1921G provides
63 bins that begin at memory address 0800h. Each bin
consists of a 16-bit, nonrolling-over binary counter that
is incremented each time a temperature value acquired
during a mission falls into the range of the bin. The least
significant byte of each bin is stored at the lower
address. Bin 0 begins at memory address 0800h, bin 1
at 0802h, and so on up to 087Ch for bin 62, as shown
in Figure 7. The number of the bin to be updated after a
temperature conversion is determined by cutting off the
two least significant bits of the binary temperature
value. Out-of-range values are range locked and count-
ed as 00h or FAh.
Since each data bin is 2 bytes, it can increment up to
65,535 times. Additional measurements for a bin that
has already reached its maximum value are not count-
ed; the bin counter remains at its maximum value. With
the fastest sample rate of one sample every minute, a
2-byte bin is sufficient for up to 45 days if all tempera-
ture readings fall into the same bin.
TEMPERATURE READING
FAh
00h
01h
02h
03h
04h
05h
06h
07h
08h
F3h
F4h
F5h
F6h
F7h
F8h
F9h
______________________________________________________________________________________
EQUIVALENT IN °C
TEMPERATURE
+85.0 or higher
-40.0 or lower
+81.5
+82.0
+82.5
+83.0
+83.5
+84.0
+84.5
-39.5
-39.0
-38.5
-38.0
-37.5
-37.0
-36.5
-36.0
HISTOGRAM BIN NUMBER
For some applications it is essential to not only record
temperature over time and the temperature histogram,
but also record when exactly the temperature exceed-
ed a predefined tolerance band and for how long the
temperature stayed outside the desirable range. The
DS1921G can log high and low durations. The toler-
ance band is specified by means of the Temperature
Alarm Threshold registers, addresses 20Bh and 20Ch
in the register page. One can set a high temperature
and low temperature threshold. See the Temperature
Conversion section for the data format the temperature
has to be written in. As long as the temperature values
stay within the tolerance band (i.e., are higher than the
low threshold and lower than the high threshold), the
DS1921G does not record any temperature alarm. If the
temperature during a mission reaches or exceeds
either threshold, the DS1921G generates an alarm and
sets either the temperature high flag (THF) or the tem-
perature low flag (TLF) in the Status register (address
Thermochron iButton
60
61
61
61
61
62
62
62
0
0
0
0
1
1
1
1
2
Temperature Alarm Logging
HISTOGRAM BIN ADDRESS
87Ah to 87Bh
87Ah to 87Bh
87Ah to 87Bh
87Ah to 87Bh
87Ch to 87Dh
87Ch to 87Dh
87Ch to 87Dh
800h to 801h
800h to 801h
800h to 801h
800h to 801h
802h to 803h
802h to 803h
802h to 803h
802h to 803h
804h to 805h
878h to 879h
17