EDS6-120 FUJI [Fuji Electric], EDS6-120 Datasheet - Page 3

no-image

EDS6-120

Manufacturer Part Number
EDS6-120
Description
E SERIESE ELECTROMAGNETIC FLOWMETER(SEPARATE TYPE)
Manufacturer
FUJI [Fuji Electric]
Datasheet
Length of straight pipe for installing the electromagnetic flowmeter
The length of the up-stream/down-stream straight pipe of the flowmeter should be long enough to ensure accurate measure-
ments. See below.
Note 1) D=diameter of measuring pipe.
Note 2) Do not put any objects, which affect magnetic field, electromotive force and flow velocity profile, in the measuring pipe.
Note 3) Use a straight pipe (2D to 3D) on the down-steam side, if the drift to the up-stream side is affected by installing valves,
etc.
1-1) Upstream reducer pipe
1-2) Upstream diffuser pipe
1-3) Various upstream valves
1-4) Upstream 90 bend
1-5) Upstream T-pipe
direction
Flowing
direction
Flowing
Minimum length of straight pipe between up-stream/down-stream joints and flowmeter.
Note) "5D" with sluice valve fully open
Flowing direction
Flowing direction
Flowing direction
Note)
5D
10D
5D
5D
5D
Flowmeter
Flowmeter
Flowmeter
Flowmeter
Flowmeter
Flowing direction
Flowing direction
2-1) Downstream reducer pipe
2-2) Downstream diffuser pipe
2-3) Various downstream valves
2-4) Downstream 90 bend
2-5) Downstream T-pipe
Flowing direction
Flowing direction
Flowing direction
Downstream straight pipe unnecessary
Downstream straight pipe unnecessary
Flowmeter
Flowmeter
Flowmeter
Flowmeter
Flowmeter
2
2
2
3D
3D
3D
3

Related parts for EDS6-120