MDO4054-3GSA Tektronix, MDO4054-3GSA Datasheet - Page 15

no-image

MDO4054-3GSA

Manufacturer Part Number
MDO4054-3GSA
Description
Specifications: Scope Type: Bench ; Scope Channels: 4 Analogue + 16 Digital + 1 RF ; Bandwidth: 500MHz ; Meter Display Type: XGA TFT-LCD Color ; Sampling Rate: 2.5GSPS ; Input Impedance: 1Mohm ; Rise Time: 700ps ; Input Voltage: 300Vrms ; Supply Volt
Manufacturer
Tektronix
Datasheet
*
Acquisition Modes
Mode
Sample
Peak Detect
Averaging
Envelope
Hi Res
Roll
Mode
MIL-STD-1553
(Optional)
I
(Optional)
Parallel
1
2
S/LJ/RJ/TDM
Trigger selection of Command Word will trigger on Command and ambiguous Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous Command/Status words.
Description
Trigger on Sync, Word Type*
and Parity individually), Status Word*
Subsystem Flag, Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity individually), Data Word (user-specified 16-bit data value),
Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time (minimum time selectable from 4 µs to 100 µs; maximum time selectable from
12 µs to 100 µs; trigger on < minimum, > maximum, inside range, outside range). RT Address can be further specified to trigger on =, ≠, <, >, ≤,
≥ a particular value, or inside or outside of a range.
Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific data value, or inside or outside of
a range
Maximum data rate for I
Maximum data rate for TDM is 25 Mb/s
Trigger on a parallel bus data value. Parallel bus can be from 1 to 20 bits in size. Binary and Hex radices are supported
Description
Acquire sampled values
Captures glitches as narrow as 800 ps (1 GHz
models) or 1.6 ns (500 MHz models) at all sweep
speeds
From 2 to 512 waveforms included in average
Min-Max envelope reflecting Peak Detect data over
multiple acquisitions
Real-time boxcar averaging reduces random noise
and increases vertical resolution
Scrolls waveforms right to left across the screen at
sweep speeds slower than or equal to 40 ms/div
2
S/LJ/RJ is 12.5 Mb/s
1
(Command, Status, Data), Command Word*
1
(set RT Address, Message Error, Instrumentation, Service Request Bit, Broadcast Command Received, Busy,
Waveform Measurements
Measurement
Cursors
Automatic
Measurements –
Time Domain
Automatic
Measurements –
Frequency Domain
Measurement
Statistics
Reference Levels
Gating
Waveform Histogram
Waveform Histogram
Measurements
1
(set RT Address, T/R, Sub-address/Mode, Data Word Count/Mode Code,
Mixed Domain Oscilloscopes — MDO4000 Series
Description
Waveform and Screen
29, of which up to eight can be displayed
on-screen at any one time. Measurements include:
Period, Frequency, Delay, Rise Time, Fall Time,
Positive Duty Cycle, Negative Duty Cycle, Positive
Pulse Width, Negative Pulse Width, Burst Width,
Phase, Positive Overshoot, Negative Overshoot,
Peak-to-Peak, Amplitude, High, Low, Max, Min,
Mean, Cycle Mean, RMS, Cycle RMS, Positive
Pulse Count, Negative Pulse Count, Rising Edge
Count, Falling Edge Count, Area and Cycle Area
3, of which one can be displayed on-screen at
any one time. Measurements include Channel
Power, Adjacent Channel Power Ratio (ACPR),
and Occupied Bandwidth (OBW)
Mean, Min, Max, Standard Deviation
User-definable reference levels for automatic
measurements can be specified in either percent
or units
Isolate the specific occurrence within an acquisition
to take measurements on, using either the screen
or waveform cursors
A waveform histogram provides an array of data
values representing the total number of hits
inside of a user-defined region of the display. A
waveform histogram is both a visual graph of the
hit distribution as well as a numeric array of values
that can be measured.
Sources – Channel 1, Channel 2, Channel 3,
Channel 4, Ref 1, Ref 2, Ref 3, Ref 4, Math
Types – Vertical, Horizontal
Waveform Count, Hits in Box, Peak Hits, Median,
Max, Min, Peak-to-Peak, Mean, Standard
Deviation, Sigma 1, Sigma 2, Sigma 3
www.tektronix.com
15

Related parts for MDO4054-3GSA