U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 64

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
2
46
Operations and Features
Percentage (%)
This operation allows you to transfer the measurement value with a
proportional percentage display. You can set the upper limit (HI) and
lower limit (LO) to any value between 0 and ±100% of the highest range
for the present function.
The percentage operation calculates and display the measurement value
using the equation below:
Percentage (%) = [Measurement value – LO/(HI- LO)] x 100%]
When the HI limit is equal to LO limit, the percentage operation will use
the equation as below:
Percentage (%) = [(Measurement value – HI)/HI] x 100%]
Figure 2-17 Typical percentage (%) operation display
When enabled, the actual readings are shown in primary display and the
percentage’s result is shown in secondary display.
• The secondary display shows OL when readings exceed the maximum
• When autoranging is enabled, this operation will be used to lock the
display of 999.99%.
current range.
U3401A User’s and Service Guide