2
Table 2-3 Setup menu and communications parameters
34
First tier
menu
rS232
Ref Ω
bEEP
rHold
[1]
[2]
setting value.
For calibration use only.
Variation count is used to recognize new value will be updated once the variation of measuring value is esceeded the
Operations and Features
[1]
Description
RS232
interface
parameters
Reference
impedance
Beeper
selection
Refresh hold
Using the Setup Menu
The Setup menu allows you to customize a number of non- volatile
instrument configurations. The content of the Setup menu are shown in
Table 2-
Second
tier menu
bAUd
PArtY
dAtA
StoP
Echo
Print
bEEP
bEEPF
StAtE
Count
3.
Description
Baud rate for remote
communication with a PC
(remote control).
Parity bit for remote
communication with a PC.
Data bit length for remote
communication with a PC.
Stop bit.
ECHO. Return a character to PC
in remote communication.
Printer-Only. Print measured
data to a PC in remote
communication.
Enable or disable beep
operation.
Beep frequency.
State.
Variation count.
[2]
Default
factory
setting
9600
None
8
1
OFF
OFF
600 Ω
ON
4096 Hz
OFF
300
U3401A User’s and Service Guide
Available settings
9600, 4800, 2400, 1200,
600 and 300
None, odd or even
7 or 8
1 or 2
ON or OFF
ON or OFF
8000 Ω, 1200 Ω, 1000 Ω,
900 Ω, 800 Ω, 600 Ω,
500 Ω, 300 Ω, 250 Ω,
150 Ω, 135 Ω, 125 Ω,
124 Ω, 110 Ω, 93 Ω, 75 Ω,
50 Ω, 16 Ω, 8 Ω, 4 Ω, 2 Ω
ON or OFF
4096, 2048, 1024 8192 Hz
ON or OFF
100, 200, 300, 400, 500,
600, 700, 800, 900, 1000.