U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 42

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
Figure 2-4 ACI RMS or DCI (10A) terminal connection and display
2
24
Operations and Features
Measuring AC(RMS) or DC Current up to 10 A
1 Press
2 Power off the measured circuit.
3 Connect the red and black test leads to the A input terminal as shown
4 Probe the test points in series with the circuit.
5 Power on the measured circuit and read the display.
• One range:
• Shunt resistance:
• Input protection:
in
current source
Figure 2-
ACI or DCI
+
4.
or
.
10.000 A for DC or AC RMS continuous
12.000 A DC or AC RMS for maximum 30 seconds
0.01
Internal 25 A, 440 V fuse for 10A terminal
to 100 Ω for 500 μA to 10 A ranges