72-820 TENMA, 72-820 Datasheet - Page 19

INFRARED THERMOMETER, 18°C TO 550°C

72-820

Manufacturer Part Number
72-820
Description
INFRARED THERMOMETER, 18°C TO 550°C
Manufacturer
TENMA
Datasheet

Specifications of 72-820

Temperature Measuring Range
18°C To 550°C
Temperature Tester Type
Thermometer - Infrared
Accuracy %
1.8%
Wavelength Max
670mm
Model 72-820: OPERATING MANUAL
6.
7.
1.
2.
buss connections.
Press MODE to select MAX.
Scan each end cap on each fuse/
Press SET and then press
relatively low for uninsulated connectors or buss
connections or relatively high for insulated
connections.
Scan the conductor, moving toward direction of
electrical connector (quick connect, wire nut, buss
connection, or lug).
/
to set emissivity to
1.
2.
3.
4.
5.
6.
Turn off heating, cooling, and blower.
Press SET to select emissivity. Press
emissivity relatively high for painted surfaces or
window surfaces.
Press MODE to select MIN when opposite side of
wall is at lower temperature and or select MAX when
opposite side of wall is at higher temperature.
Measure an interior partition wall surface temperature.
Do not release the trigger. Record this temperature
as your baseline (or benchmark) for a “perfectly”
insulated wall.
Face the wall to be scanned. Stand 1.5m away to
scan a 5cm spot on the wall.
Scan horizontal rows of wall from top to bottom, or
horizontal rows of ceiling from wall to wall. Look for
greatest deviations from baseline temperature to
identify problems. This completes the insulation test
scan.
/
to select
17